Housings<br />

164<br />

Retractable Housings<br />

Sensor Access Without Process Interruption<br />

Mettler-Toledo Ingold International<br />

InTrac 798e<br />

The Hygienic Solution<br />

Feature Overview<br />

<strong>–</strong> Standard 120 mm sensors fit<br />

all <strong>housings</strong>: more cost effective<br />

<strong>–</strong> Unique self-contained design<br />

with no external moving parts:<br />

very safe operation<br />

<strong>–</strong> EHEDG compliance<br />

Other Highlights<br />

<strong>–</strong> 3.1 certificate included<br />

<strong>–</strong> Optimized drain placement assures<br />

no residual remains after cleaning<br />

<strong>–</strong> Powerful non-jamming mechanism<br />

works well in viscous or fibrous environments<br />

®<br />

Originally designed as a companion product to protect the InPro 3300 ISFET<br />

pH electrode from caustic CIP conditions, the unique benefits of the InTrac 798e<br />

housing can also be used with all 120 mm sensors. The InTrac 798e is<br />

completely self-contained and has none of the external moving parts commonly<br />

associated with <strong>retractable</strong> <strong>housings</strong>. Regardless of the insertion depth selected,<br />

less costly 120 mm sensors are needed to operate the InTrac 798e, which<br />

reduces overall operating expenses and inventory requirements.<br />

Absolute cleanability is possible because the flushing chamber drain and O-ring<br />

have been optimally positioned to assure all CIP solutions are completely drained<br />

from the chamber prior to the sensor being re-inserted into your <strong>process</strong>.<br />

Specifications<br />

Wetted parts Stainless 316L<br />

Surface finish N5/N5 (Ra16/Ra16)<br />

(O-ring groove/Other)<br />

O-ring EPDM-FDA<br />

Sensor fitting Pg13.5<br />

Temperature range 0<strong>–</strong>140°C/32<strong>–</strong>284°F<br />

Pressure rating 0<strong>–</strong>16 bar (0<strong>–</strong>232 psi)<br />

(Sensor dependent)<br />

Process Entry: Top-entry, Side-entry, Pipe-entry (see Figure 1 pg. 142)<br />

Suggested Sensors<br />

pH/DO/CO2/Conductivity<br />

All 120 mm sensors with Pg13.5 and 12 mm diameter (except turbidity sensors)<br />

Sensor Fit Guide<br />

Sensor Length Insertion Length<br />

50 mm 100 mm<br />

120 mm • •<br />

A unique design allows the InTrac 798e housing to use 120 mm sensors regardless of the<br />

housing length.