Appendix Babylog 8000<br />

12.1.3 Amplitude setting <strong>and</strong> oscillatory volumes<br />

The relationship between the relative oscillation amplitude <strong>and</strong> the<br />

resulting volume is non-linear. The following illustration shows that<br />

only below 65% the oscillatory volume goes down. This was measured<br />

at 25 mbar MAP (patient circuit <strong>and</strong> humidifier: Fisher-Paykel; circuit<br />

compliance: 0.77 ml/mbar; test lung compliance: 0.66 ml/mbar).<br />

Vt-hf<br />

set Amplitude<br />

<strong>Frequency</strong><br />

Figure 12.3: Oscillatory volume as a function of amplitude <strong>and</strong> frequency<br />

setting<br />

51