21.06.2018 Views

Noris Medical Dental Implants Product Catalog 2018 3

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Mono<br />

L<br />

D1<br />

D<br />

Mono implants are specifically indicated for replacing<br />

maxillary lateral incisors and mandibular central and<br />

lateral incisors.<br />

The implants are cleared for immediate, non-occlusal<br />

provisionalization in single-tooth restorations.<br />

Multiple-unit restorations should be splinted together.<br />

In appropriate clinical conditions MONO <strong>Implants</strong><br />

may be loaded immediately.<br />

Material: Titanium (Ti6Al4V ELI)<br />

Treatment: RBM<br />

ONE-PIECE IMPLANT<br />

L 3 6.5<br />

2.15<br />

D2<br />

D1<br />

D<br />

2.15<br />

Drilling Speed (RPM)<br />

1200-<br />

1500<br />

1200-<br />

1500<br />

200-<br />

500<br />

900-<br />

1200<br />

200-<br />

500<br />

500-<br />

700<br />

200-<br />

500 400-700<br />

200-<br />

500<br />

Drill Diameter md ø1.5 ø2.0 ø2.0 ø2.8 ø2.8 ø3.2 ø3.2 ø3.65<br />

Drill Sequence*<br />

Bone type D1, D2<br />

6<br />

10<br />

16<br />

ø3.0 ø3.3 ø3.75 ø4.2 ø5.0<br />

Drilling Speed (RPM)<br />

1200-<br />

1500<br />

1200-<br />

1500<br />

200-<br />

500<br />

900-<br />

1200<br />

500-<br />

700<br />

500-<br />

700<br />

Drill Diameter md ø1.5 ø2.0 ø2.0 ø2.8 ø3.2<br />

Drill Sequence*<br />

Bone type D3, D4<br />

6<br />

10<br />

16<br />

ø3.0 ø3.3 ø3.75 ø4.2 ø5.0<br />

* The proposed procedure is only a recommendation and should not replace the doctor's judgment.<br />

The implants may be placed in immediate function when good primary stability (above 35 Ncm) has been achieved and with appropriate occlusal loading.<br />

/ 28 / <strong>Implants</strong>

Hooray! Your file is uploaded and ready to be published.