Surgeon’s Preset Center<br />
HF-GENERATOR<br />
A preset center allows up to 10 surgeons to store their preferred power settings<br />
in the Cut, Coagulation and Bipolar Modes.<br />
Cutting<br />
A digitally controlled and regulated cutting effect. The intelligent digital system<br />
technology with FAST DIGITAL FEETBACK SYSTEM (FDFS TM<br />
) permits reproducible<br />
cuts regardless of the tissue type. FDFS TM<br />
technology offers digitally controlled<br />
cutting effects for efficient operation and optimal results with a high<br />
degree of safety. This innovative cut behaviour can be utilized in a variety of procedures,<br />
from high-powered/high-impedance procedures (CUT I) to general<br />
applications (CUT II).<br />
CUT I mode has a digitally controlled cutting quality that ensures performance<br />
when cutting adipose tissue or cutting underwater with minimal output power.<br />
CUT II mode is a less aggressive cutting effect with the highest degree of safety<br />
to precise effects in general applications or microsurgery.<br />
Blended cutting mode with 10 blend settings affords the most effective blend setting<br />
for the desired effect during any kind of procedure.<br />
Calibration Free<br />
The Promed <strong>PB</strong>-<strong>400</strong> does not need any calibration due to its digital technology.<br />
Coagulation<br />
A digital controlled and regulated coagulation effect.<br />
Promed <strong>PB</strong>-<strong>400</strong> coagulation technology performes monopolar and bipolar coagultion<br />
effectively in every type of tissue and a variety of procedures. Pinpoint<br />
coagulation mode delivers a digitally controlled rapid desiccation effect with<br />
minimally carbonization and less sticking of the active electrode. Spray coagulation<br />
mode gives digitally controlled non-contact coagulation with a slight penetration<br />
of the tissue surface. Bipolar mode also uses digital technology to reduce<br />
instrument sticking and minimize tissue destruction.<br />
Foot Switch Indicator<br />
A display on the front of the generator clearly indicates if one or two<br />
foot switches are connected.<br />
Plug and Play<br />
ELECTROSURGICAL GENERATOR<br />
<strong>PB</strong>-<strong>400</strong><br />
The <strong>PB</strong>-<strong>400</strong> is ready to use: plug in the instrument – start to work.<br />
Order-No.: 670-91011<br />
Calibration Free<br />
The promed <strong>PB</strong>-<strong>400</strong> does not need any<br />
calibration due to its digital technology.<br />
Mode Output Power<br />
Cut I <strong>400</strong> W@300 Ω<br />
Cut II 300 W@300 Ω<br />
Blend (Max) 200 W@300 Ω<br />
Pinpoint 120 W@500 Ω<br />
Spray 80 W@500 Ω<br />
Bipolar Max. 120 W@150 Ω<br />
Surgeon’s Preset Center<br />
Cutting<br />
Calibration Free<br />
Coagulation<br />
European Representative A Professional in <strong>Medical</strong> <strong>Equipment</strong> & Device
670-91011<br />
<strong>MGB</strong>/285 ProdBlatt <strong>PB</strong>-<strong>400</strong> ■ <strong>PB</strong>-<strong>400</strong> 09.07.2007<br />
Specifications may change without prior notice.<br />
GENERAL<br />
Order-No.<br />
670-91015<br />
670-91020<br />
Main Connection<br />
Power supply<br />
Frequency<br />
Classification<br />
Approval<br />
Dimension and Weight<br />
Dimension (w/h/d)<br />
Weight<br />
Operating Conditions<br />
Operating temperature<br />
Operating humidity<br />
Extent of Delivery<br />
<strong>PB</strong>-<strong>400</strong><br />
Interface<br />
Monopolar connection<br />
Bipolar connection<br />
Ground pad cable connection (GC2HF6)<br />
TECHNICS<br />
<strong>PB</strong>-<strong>400</strong><br />
Foot Switch bipolar<br />
Foot Switch monopolar (double pedal)<br />
100-240 V AL<br />
50/60 Hz<br />
311 x 153 x 413 mm<br />
8.75 kg<br />
+10°C – +40°C<br />
30 % – 75 %<br />
■ Unit<br />
■ Main power cable<br />
■ User manual<br />
<strong>MGB</strong> Endoskopische Geräte GmbH Berlin<br />
Schwarzschildstrasse 6 · 12489 Berlin, Germany<br />
Phone: +49-30 -63 92 -7010/- 7000 · Fax: +49- 30-63 92 - 7011<br />
E-mail: sales @ mgb- berlin.de · Internet: www.mgb- berlin.de<br />
SPECIFICATION<br />
HF-Specification<br />
Cut I<br />
Cut II<br />
Blend (Max)<br />
Pinpoint<br />
Spray<br />
Bipolar<br />
<strong>400</strong> W at 300 Ω<br />
300 W at 300 Ω<br />
200 W at 300 Ω<br />
120 W at 500 Ω<br />
80 W at 500 Ω<br />
Max. 120 W at 150 Ω<br />
Order-No. Standard Accessories<br />
PR310P5 (1 box/10 pcs.)<br />
EB22R2 (1 box/50 pcs.)<br />
EN22R2 (1 box/50 pcs.)<br />
ES22R3 (1 box/50 pcs.)<br />
GB900 (1 box/25 pcs.)<br />
GC2VN (Ground pad cable connection)<br />
(1 box/50 pcs.)<br />
FAG 1221 (See Promed catalogue)<br />
FC813FS<br />
Single Foot Switch<br />
A Professional in <strong>Medical</strong> <strong>Equipment</strong> & Device<br />
■ Reusable ES pencil, pencil type: PR 310<br />
■ Electrode: standard Blade<br />
■ Cable length: 5 m,<br />
■ Plug type: 3-prong plug<br />
■ Reusable electrode, standard length<br />
■ Tip type: Blade, insulation overmold<br />
■ Reusable electrode, standard length<br />
■ Tip type:needle, insulation overmold<br />
■ Reusable electrode, standard length<br />
■ Tip type: 3 mm ball, insulation overmold<br />
■ Ground pad, bipolar<br />
■ Pad size: adults<br />
■ Cable type: NA<br />
■ Plug: NA<br />
■ Plug: Vally cab type<br />
■ Reusable bipolar forceps<br />
■ Reusable 5m Silicone cable,<br />
■ 2 x 4 Banana plug,<br />
■ Bipolar forceps flat connector<br />
Double Foot Switch