You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
www.rego.de<br />
<strong>REGO</strong> Testing and Disposal<br />
of X-Ray Protection Aprons<br />
Testing of Aprons<br />
In addition to the daily optical test of the aprons by checking tears, holes, defective seams and possibly<br />
defects of the inner protection material, a periodical touch or x-ray check is necessary (DIN 6857-2), so small<br />
damages can be detected. Depending on the age and frequency of use of the X-ray protection apron, a regular<br />
and at least annual touch test is recommended according to DIN 6857-2. Noticeable damage such as breaks,<br />
kinks, cracks or holes can be hereby detected.<br />
Latest after the third year of use the aprons must be tested by X-Ray. Only this way damages can be recognized<br />
and removed. When there was no damage, the next test will have to be done after 2 years, and then every<br />
year.<br />
We offer a professional testing service. Send your X-Ray protection clothes to us, we check them and return<br />
them to you at once. The service includes the regular sight and functional testing – depending on the age –<br />
as well as the testing with X-Rays. The necessary documentation is provided as well.<br />
Please contact us for more information.<br />
Disposal of Aprons<br />
Once put in the trash, is now a valuable ressource for the industry through sustainable recycling. With this<br />
motivation we made it our duty, as experienced manufacturer of X-Ray protection apparell, to contribute as<br />
much as possible for the proper disposal, in order to reach the goal of a more efficient recycling economy.<br />
We highly value the correct disposal of our products and are also taking care of the professional recycling.<br />
X-Ray Protection Apparell has to be disposed as hazardrous waste due to ist toxicity. Therefore this process<br />
is better left to professionals.<br />
We would be very pleased to take over the disposal of your x-ray protection gear.<br />
40