18.01.2013 Views

Viasys Healthcare Infant Flow - SiPAP paediatric respiratory support

Viasys Healthcare Infant Flow - SiPAP paediatric respiratory support

Viasys Healthcare Infant Flow - SiPAP paediatric respiratory support

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

SAFETY ACTION NOTICE<br />

By arrangement with the Scottish Government Health Directorates<br />

VIASYS HEALTHCARE INFANT FLOW ® SIPAP<br />

PAEDIATRIC RESPIRATORY SUPPORT DEVICE:<br />

CHANGES TO OPERATOR MANUAL<br />

SUMMARY<br />

SAN(SC)07/56<br />

11 DEC 2007<br />

Medical Devices<br />

Page 1 of 2 Pages<br />

The operator manual of <strong>Viasys</strong> <strong>Healthcare</strong>’s <strong>Infant</strong> <strong>Flow</strong> ® <strong>SiPAP</strong> <strong>paediatric</strong> <strong>respiratory</strong> <strong>support</strong><br />

device has been updated. All relevant staff should be made aware of this change.<br />

BACKGROUND<br />

1. The <strong>Infant</strong> <strong>Flow</strong> ® <strong>SiPAP</strong> <strong>paediatric</strong> <strong>respiratory</strong> <strong>support</strong> device, manufactured by <strong>Viasys</strong><br />

<strong>Healthcare</strong>, provides a non-invasive form of <strong>respiratory</strong> <strong>support</strong> designed for infants in hospital<br />

environments and for transporting these patients within the hospital. <strong>Infant</strong> <strong>Flow</strong> ® <strong>SiPAP</strong><br />

provides bi-level nasal CPAP for the spontaneously breathing neonate.<br />

2. The operator manual used to state incorrectly that changes to parameters will only be set if<br />

confirmed by pressing the control button. In fact, changes are automatically accepted in the event<br />

that no other button is pressed within 15 seconds.<br />

3.<br />

The operator manual has been corrected. Revision G, page 23 ‘Changing a control’ now states:<br />

“When a control setting such as Ti (inspiratory time) is selected, increase and decrease buttons<br />

appear. The control and displayed value for the selected parameter are highlighted. Use the<br />

decrease or increase keys to adjust the parameter. Accept the action by pressing the control button<br />

again.<br />

“In normal treatment screens, parameter changes take immediate effect.<br />

“If no button is pressed for 15 seconds during this procedure, highlighted controls are<br />

automatically adjusted according to the last key stroke and the screen is restored to its previous<br />

configuration.”<br />

ACTION<br />

4. This notice should be brought to the attention of all appropriate managers, staff and users.<br />

5. A copy of the current operator manual should be obtained from the <strong>Viasys</strong> <strong>Healthcare</strong> website (see<br />

References). At the time of publishing this is Revision G dated May 2006.<br />

6. All relevant staff should be made aware of the change to the operator manual and have access to a<br />

copy of the current instructions.<br />

Suggested Distribution Accident & Emergency Anaesthetics Device Managers<br />

Health & Safety Intensive Therapy Units Maternity Medical<br />

Medical Physics Neonatal Nursing Operating Departments<br />

Paediatrics Resuscitation Teams Risk Management Special Care Baby Units<br />

SCOTTISH HEALTHCARE SUPPLIES<br />

Gyle Square Edinburgh EH12 9EB<br />

A Division of the National Service Scotland for NHSScotland<br />

CONTACT EMAIL: iric@shs.csa.scot.nhs.uk FAX: 0131 314 0722<br />

WEBSITE: http://www.nhsscotland.com/shs/hazards_safety/adverse_p.html N:\WORD\HAZ\IRIC PUBLICATIONS\FINAL SAN\PUBLIC SAN\PSAN0756.DOC


SAFETY ACTION NOTICE<br />

By arrangement with the Scottish Government Health Directorates<br />

VIASYS HEALTHCARE INFANT FLOW ® SIPAP<br />

PAEDIATRIC RESPIRATORY SUPPORT DEVICE:<br />

CHANGES TO OPERATOR MANUAL<br />

ENQUIRIES<br />

Enquiries to the manufacturer should be address to:<br />

Mr Detlef Grotheer<br />

<strong>Viasys</strong> <strong>Healthcare</strong> GmbH<br />

7 Leibnizstrasse<br />

97204 Hoechberg<br />

Germany<br />

Tel: 0049 931 4972 111<br />

E-mail: detlef.grotheer@viasyshc.com<br />

REFERENCES<br />

1. <strong>Infant</strong> <strong>Flow</strong> ® <strong>SiPAP</strong> Operator Manual, Revision G May 2006:<br />

http://www.viasyshealthcare.com/prod_serv/downloads/197_PN_675-101-101G_<strong>Infant</strong><br />

<strong>Flow</strong>_<strong>SiPAP</strong>_OM.pdf<br />

2. Product website:<br />

SCOTTISH HEALTHCARE SUPPLIES<br />

Gyle Square Edinburgh EH12 9EB<br />

A Division of the National Service Scotland for NHSScotland<br />

SAN(SC)07/56<br />

11 DEC 2007<br />

Medical Devices<br />

Page 2 of 2 Pages<br />

http://www.viasyshealthcare.com/prod_serv/BUOverview.aspx?config=ps_appdtl&id=53<br />

CONTACT EMAIL: iric@shs.csa.scot.nhs.uk FAX: 0131 314 0722<br />

WEBSITE: http://www.nhsscotland.com/shs/hazards_safety/adverse_p.html N:\WORD\HAZ\IRIC PUBLICATIONS\FINAL SAN\PUBLIC SAN\PSAN0756.DOC

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!