<strong>QC1</strong> <strong>ESR</strong><br />

REF<br />

5920005<br />

DESCRIPCION<br />

Level 1 Normal<br />

Level 2 Abnormal<br />

Sólo para uso diagnóstico in vitro<br />

1 x 1,5 mL<br />

1 x 1,5 mL<br />

<strong>QC1</strong> <strong>ESR</strong><br />

Control de referencia de sangre total<br />

Velocidad de Sedimentación Globular<br />

FUNDAMENTO<br />

<strong>QC1</strong> ERS, , es un material de referencia empleado en el control<br />

de precisión de la Velocidad de Sedimentación Globular (VSG) en<br />

los Instrumentos de VSG de <strong>Linear</strong> Chemicals.<br />

Level 1 Normal<br />

Level 2 Abnormal<br />

COMPOSICIÓN<br />

Dosificado en tubos de plástico.<br />

Level 1 tapón verde, Level 2 tapón rojo.<br />

Hematíes humanos estabilizados.<br />

Suspendidos en un medio tamponado.<br />

Contiene bioacidas.<br />

Precauciones: Evitar el contacto con la piel y las mucosas. Los<br />

componentes de origen humano presentes en este material han dado<br />

resultados negativos a tests de detección de anticuerpos del Virus de la<br />

Inmunodeficiencia Humana (VIH), antígeno de superficie de la Hepatitis B<br />

(HBsAg) y el virus de la Hepatitis C (HCV). Sin embargo, deben manipularse<br />

como si se tratara de una muestra potencialmente infecciosa.<br />

ALMACENAMIENTO Y ESTABILIDAD<br />

Conservar a 18-30ºC.<br />

Estable hasta la fecha de caducidad indicada en la etiqueta del<br />

tubo. Siempre que permanezca bien cerrado, a fin prevenir la<br />

evaporación de los hematíes. No congelar ni exponer a calor<br />

excesivo. Evitar la exposición prolongada de los tubos abiertos a<br />

la luz.<br />

PREPARACIÓN<br />

Los tubos <strong>QC1</strong> ERS están listos para su uso.<br />

PROCEDIMIENTO<br />

<strong>QC1</strong> ERS debe ensayarse de acuerdo con las instrucciones de<br />

uso del instrumento y ser tratado como la muestra problema.<br />

Homogeneizar la sangre, manual o automáticamente, por inversión<br />

durante 2 o 5 minutos, dependiendo del tiempo que ha estado en<br />

reposo. Evitar la formación de espuma.<br />

RANGO DE VALORES<br />

Los valores indicados son específicos de cada lote y se han<br />

obtenido a partir de replicados en los Instrumentos de VSG de<br />

<strong>Linear</strong> Chemicals.<br />

Las medias de cada laboratorio deben estar comprendidas entre el<br />

correspondiente rango aceptable, pero pueden apartarse de los<br />

valores indicados durante el periodo de vida útil del producto, las<br />

variaciones a lo largo del tiempo y entre laboratorios pueden<br />

deberse a diferencias en la técnica usada y al método de<br />

calibración del instrumental.<br />

Se recomienda que cada laboratorio establezca sus propias<br />

medidas y rangos aceptables.<br />

CAUSAS DE ERROR<br />

- Material sucio (no re-usar material desechable).<br />

- Hemólisis de la muestra.<br />

- El exceso de anticoagulante provoca un descenso del valor.<br />

- La inclinación de los tubos aumenta su valor.<br />

- Burbujas de aire en las muestras.<br />

- Temperaturas inferiores a 18ºC dan resultados inferiores.<br />

(consultar Tabla de Manley).<br />

- Agitación insuficiente de los hematíes.<br />

LIMITACIONES DEL PRODUCTO<br />

<strong>Linear</strong> <strong>QC1</strong> ERS no esta diseñado para ser utilizado como control de<br />

otras pruebas hematológicas.<br />

BIBLIOGRAFIA<br />

1. Wintrobe, M.: Clinical Hematology. Lea and Fibiger, Philadelphia, pp. 314-<br />

322, (1956).<br />

2. Miale, J.: Laboratory Medicine Hematology, The C.V. Mosby Company, St.<br />

Louis, pp 469-475, (1972).<br />

3. Koepke, Bull, Gilmer and Goldblatt: Hematology in Quality Assurance<br />

Practices for Health Laboratories, Interdisciplinary Books: American Public<br />

Health Assn., (1978).<br />

<strong>QC1</strong> <strong>ESR</strong> Lot # 59143 Exp ver etiqueta del producto<br />

ERS analyzer Tubo Level 1 Normal Level 2 Abnormal<br />

THERMA NE, ERILINE AR & LENA (<strong>Linear</strong> C., Spain) 1 Plástico 7-15 mm/h 47-61 mm/h<br />

IRIA & THERMA (<strong>Linear</strong> Chemicals, Spain) Plástico 6-17 mm/h 63-87 mm/h<br />

Los resultados son aplicables a este lote y a su sublotes, indentificados con letras alfabéticas que siguen al nº de lote.<br />

B5920-3/0506<br />

R1.cas<br />

QUALITY SYSTEM CERTIFIED<br />

ISO 9001 ISO 13485<br />

LINEAR CHEMICALS S.L. Joaquim Costa 18 2ª planta. 08390 Montgat, Barcelona, SPAIN<br />

Telf. (+34) 934 694 990 Fax. (+34) 934 693 435. website www.linear.es


<strong>QC1</strong> <strong>ESR</strong><br />

REF<br />

5920005<br />

DESCRIPTION<br />

Level 1 Normal<br />

Level 2 Abnormal<br />

Only for in vitro diagnostic use<br />

1 x 1.5 mL<br />

1 x 1.5 mL<br />

<strong>QC1</strong> <strong>ESR</strong><br />

Whole blood reference control<br />

Erythrocyte Sedimentation Rate<br />

PRINCIPLE<br />

<strong>QC1</strong> ERS, is a whole blood reference control to monitor patient<br />

erythrocyte sedimentation rate (<strong>ESR</strong>) procedures on <strong>Linear</strong> ERS<br />

Instruments.<br />

Level 1 Normal<br />

Level 2 Abnormal<br />

COMPOSITION<br />

Stabilized human red cells suspended in<br />

a buffer. Contains biocides.<br />

Dispensed in plastic tubes.<br />

Level 1 green cap, Level 2 red cap.<br />

Precautions: Avoid any contact with skin or mucous membranes.<br />

The human constituents have tested negative for antibodies to<br />

Human Immunodeficiency Virus (HIV), Hepatitis B surface Antigen<br />

(HBsAg) and Hepatitis C Virus (HCV). Handle as potentially<br />

infectious.<br />

STORAGE AND STABILITY<br />

Store at 18-30ºC. Stable up to the expiration date stated on<br />

the label. Do not freeze. Do not expose to excessive heat.<br />

Avoid prolonged exposure of opened tubes to light.<br />

<strong>QC1</strong> ERS is ready-to-use.<br />

PREPARATION<br />

PROCEDURE<br />

<strong>QC1</strong> ERS product should be treated and tested as the patient<br />

specimens and run in accordance with the instructions<br />

accompanying the instrument.<br />

Mix the tube by using a mixing device or manually by complete<br />

inversion. Mixing time should be between 2 and 5 minutes<br />

depending upon the standing time of the tubes prior testing. Avoid<br />

foaming.<br />

EXPECTED VALUES<br />

The values printed in this insert were derived from replicates with<br />

<strong>Linear</strong> Chemicals ERS Instruments. And are specific for this lot of<br />

product. Individual laboratory means should fall within the<br />

corresponding acceptable range, however, laboratory means may<br />

vary from the listed values during his life. Variations over time and<br />

between laboratories may be caused by differences in laboratory<br />

techniques and instrumentation calibrating methods.<br />

It is recommended that each laboratory establish its own means<br />

and acceptable ranges.<br />

CAUSES OF ERROR<br />

- Dirty material (not re-use disposable material).<br />

- Hemolysed samples.<br />

- An excess of anticoagulant causes a reduction of the value..<br />

- Tube inclination causes an increase of the value.<br />

- Bubbles in sample.<br />

- Inferior Temperatures to 18ºC give inferior results.<br />

(see Manley table).<br />

- Incomplete mixing.<br />

LIMITATIONS OF THE PRODUCT<br />

<strong>QC1</strong> ERS control is not designed to be used as a control for any<br />

hematology procedures other than Erythrocyte Sedimentation Rate.<br />

REFERENCES<br />

1. Wintrobe, M.: Clinical Hematology. Lea and Fibiger, Philadelphia, pp. 314-322, (1956).<br />

2. Miale, J.: Laboratory Medicine Hematology, The C.V. Mosby Company, St. Louis, pp<br />

469-475, (1972).<br />

3. Koepke, Bull, Gilmer and Goldblatt: Hematology in Quality Assurance Practices for<br />

Health Laboratories, Interdisciplinary Books: American Public Health Assn., (1978).<br />

<strong>QC1</strong> <strong>ESR</strong> Lot # 59143 Expiry date: see product label<br />

ERS analyzer Tube Level 1 Normal Level 2 Abnormal<br />

THERMA NE, ERILINE AR & LENA (<strong>Linear</strong> C., Spain) 1 Plastic 7-15 mm/h 47-61 mm/h<br />

IRIA & THERMA (<strong>Linear</strong> Chemicals, Spain) Plastic 6-17 mm/h 63-87 mm/h<br />

These results are applicable to this lot and subsequent sublots, for which a capital letter has been assigned after<br />

the lot number.<br />

B5920-3/0506<br />

R1.ing<br />

QUALITY SYSTEM CERTIFIED<br />

ISO 9001 ISO 13485<br />

LINEAR CHEMICALS S.L. Joaquim Costa 18 2ª planta. 08390 Montgat, Barcelona, SPAIN<br />

Telf. (+34) 934 694 990 Fax. (+34) 934 693 435. website www.linear.es