Factory Calibration Certificate<br />

Example<br />

Test object: Temperature Adjustment Set (<strong>YTM03MA</strong>)<br />

Test equipment number: 0011901617<br />

Record number: 26/00<br />

Test method: Comparative temperature measurement<br />

Measuring points: 80°C, 140°C, 200°C<br />

Reference TMD Display Measurement Room<br />

TMD Correction Uncertainty Temperature<br />

[°C] [°C] [°C] [°C] [°C]<br />

80.69 80.44 0.25 0.50 23.3<br />

140.47 140.58 0.11 0.50 24.6<br />

202.03 201.39 0.64 0.50 25.8<br />

Calibration <strong>de</strong>vices used: Reference temperature measuring disk (TMD),<br />

test equipment number: TEM0100<br />

Keithley measuring instrument,<br />

test equipment number: DVM0148<br />

Traceable to: - (Physikalisch-Technische Bun<strong>de</strong>sanstalt (PTB)<br />

German Fe<strong>de</strong>ral Institute of Physics and Metrology<br />

- German Calibration Service (DKD),<br />

calibration laboratory for temperature measuring equipment<br />

Calibration Certificate for the reference TMD:<br />

- Temperature calibration for the measuring points: 80°C, 140°C, 200°C, 250°C<br />

- PTB emission rate calibration at the temperatures 80°C, 140°C and 200°C<br />

within the wavelength range of 2.5...20 µm<br />

- PTB reflectance calibration at room temperature within the wavelength range<br />

of 0.7...10 µm<br />

Recommen<strong>de</strong>d calibration interval: 1 year Next calibration: 07/01<br />

Inspector: ______________ Date: July 19, 2000<br />

Diedrich<br />