Factory Calibration Certificate<br />
Example<br />
Test object: Temperature Adjustment Set (<strong>YTM03MA</strong>)<br />
Test equipment number: 0011901617<br />
Record number: 26/00<br />
Test method: Comparative temperature measurement<br />
Measuring points: 80°C, 140°C, 200°C<br />
Reference TMD Display Measurement Room<br />
TMD Correction Uncertainty Temperature<br />
[°C] [°C] [°C] [°C] [°C]<br />
80.69 80.44 0.25 0.50 23.3<br />
140.47 140.58 0.11 0.50 24.6<br />
202.03 201.39 0.64 0.50 25.8<br />
Calibration <strong>de</strong>vices used: Reference temperature measuring disk (TMD),<br />
test equipment number: TEM0100<br />
Keithley measuring instrument,<br />
test equipment number: DVM0148<br />
Traceable to: - (Physikalisch-Technische Bun<strong>de</strong>sanstalt (PTB)<br />
German Fe<strong>de</strong>ral Institute of Physics and Metrology<br />
- German Calibration Service (DKD),<br />
calibration laboratory for temperature measuring equipment<br />
Calibration Certificate for the reference TMD:<br />
- Temperature calibration for the measuring points: 80°C, 140°C, 200°C, 250°C<br />
- PTB emission rate calibration at the temperatures 80°C, 140°C and 200°C<br />
within the wavelength range of 2.5...20 µm<br />
- PTB reflectance calibration at room temperature within the wavelength range<br />
of 0.7...10 µm<br />
Recommen<strong>de</strong>d calibration interval: 1 year Next calibration: 07/01<br />