It is possible to delay the set point checking for a<br />

programmable delay time.<br />

The PROGRAMMABLE SET POINT signal has the<br />

following default settings:<br />

− reference quantity: frequency<br />

− thereshold: 867 Hz<br />

− hysteresis: 2%<br />

− activation type: high level<br />

− delay time: 0 second<br />

NOTE<br />

The Navigator Software (optional) allows the operator<br />

to set all the programmable feature.<br />

How to connect the open collector input of the<br />

controller<br />

Here below there are the typical connections of the<br />

open collector input of <strong>TPS</strong>-compact to an external<br />

system. Two cases are considered:<br />

1. the customer supplies the 24 Vdc<br />

2. the customer does not supply the 24 Vdc.<br />

Please note that on the connector a 24 Vdc, 30<br />

mA voltage, a GROUND signal and the open collector<br />

pin are available.<br />

Case 1<br />

TECHNICAL INFORMATION<br />

Case 2 with relay utilisation<br />

Case 2 with transistor utilisation<br />