What is what<br />

1<br />

2<br />

7<br />

X-<strong>am</strong> <strong>5000</strong><br />

2<br />

3<br />

0<br />

6 4<br />

1 Gas entry 8 IR interface<br />

2 Alarm LED 9 Fastening clip<br />

3 Buzzer 10 Type plate<br />

4 [OK] key 11 Charging contacts<br />

5 Power pack 12 Measured gas display<br />

6 [ + ] key 13 Measured value display<br />

7 Display 14 Special symbols<br />

15 Tool for replacing the sensor<br />

Special symbols:<br />

Fault message<br />

Warning<br />

Display peak value<br />

Show TWA<br />

Show STEL<br />

Bump test mode<br />

Fresh air calibration<br />

5<br />

15<br />

8<br />

9<br />

10<br />

11<br />

ch4<br />

%LEL<br />

O2<br />

Vol%<br />

CO<br />

ppm<br />

H2S<br />

ppm<br />

NH3<br />

ppm<br />

12 13 14<br />

1-button calibration<br />

Single gas calibration<br />

Password required<br />

Battery 100 % full<br />

Battery 2/3 full<br />

Battery 1/3 full<br />

Battery empty<br />

00123896.eps<br />

Configuration<br />

For the individual configuration of an instrument with standard configuration, the<br />

instrument must be connected to a PC via the USB infrared cable (order no.<br />

83 17 409) or the E-Cal system. The PC software "Dräger CC-Vision" is used to<br />

perform the configuration.<br />

– Changing the configuration: see Technical Manual.<br />

Standard instrument configuration:<br />

Dräger X-<strong>am</strong> <strong>5000</strong><br />

Bump test mode 1)<br />

Fresh air calibr. 1)<br />

Life sign 1)<br />

Switch off 1<br />

LEL factor 1)<br />

(ch 4 )<br />

Averaging time 1)<br />

1) Different settings can be selected to meet customer requirements on delivery. The current setting<br />

can be checked and changed with the Dräger CC Vision software.<br />

First commissioning<br />

Quick bump test<br />

On<br />

On<br />

allowed<br />

4.4 (vol. %)<br />

(4.4 vol. % corresponds to 100 %LEL)<br />

15 minutes for STEL<br />

8 hours for TWA<br />

Before using the instrument for the first time, insert the batteries supplied or<br />

a charged T4 NiMH power pack T4 (HBT 0000 type, order no. 83 18 704) /<br />

T4 HC (HBT 0100 type, order no. 83 22 244) (refer to chapter "Replacing the<br />

batteries" on Page 15). The Dräger X-<strong>am</strong> <strong>5000</strong> is ready for operation.<br />

Operation<br />

Switching on the instrument<br />

• Press and hold the [OK] key for approx. 3 seconds until the countdown<br />

»3.2.1« shown on the display has elapsed.<br />

– All the display segments, including the visual, audible and vibration alarms,<br />

are activated for a short time.<br />

– The software version is displayed.<br />

– The instrument performs a self test.<br />

– The next sensor to be calibrated is displayed, along with the remaining days<br />

to the next calibration, e. g. » Ex%LEL CAL20«.<br />

– The time until the bump test interval elapses is displayed in days, e.g. » bt 123 «.<br />

12