Dräger X-am 5000
Dräger X-am 5000
Dräger X-am 5000
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
What is what<br />
1<br />
2<br />
7<br />
X-<strong>am</strong> <strong>5000</strong><br />
2<br />
3<br />
0<br />
6 4<br />
1 Gas entry 8 IR interface<br />
2 Alarm LED 9 Fastening clip<br />
3 Buzzer 10 Type plate<br />
4 [OK] key 11 Charging contacts<br />
5 Power pack 12 Measured gas display<br />
6 [ + ] key 13 Measured value display<br />
7 Display 14 Special symbols<br />
15 Tool for replacing the sensor<br />
Special symbols:<br />
Fault message<br />
Warning<br />
Display peak value<br />
Show TWA<br />
Show STEL<br />
Bump test mode<br />
Fresh air calibration<br />
5<br />
15<br />
8<br />
9<br />
10<br />
11<br />
ch4<br />
%LEL<br />
O2<br />
Vol%<br />
CO<br />
ppm<br />
H2S<br />
ppm<br />
NH3<br />
ppm<br />
12 13 14<br />
1-button calibration<br />
Single gas calibration<br />
Password required<br />
Battery 100 % full<br />
Battery 2/3 full<br />
Battery 1/3 full<br />
Battery empty<br />
00123896.eps<br />
Configuration<br />
For the individual configuration of an instrument with standard configuration, the<br />
instrument must be connected to a PC via the USB infrared cable (order no.<br />
83 17 409) or the E-Cal system. The PC software "Dräger CC-Vision" is used to<br />
perform the configuration.<br />
– Changing the configuration: see Technical Manual.<br />
Standard instrument configuration:<br />
Dräger X-<strong>am</strong> <strong>5000</strong><br />
Bump test mode 1)<br />
Fresh air calibr. 1)<br />
Life sign 1)<br />
Switch off 1<br />
LEL factor 1)<br />
(ch 4 )<br />
Averaging time 1)<br />
1) Different settings can be selected to meet customer requirements on delivery. The current setting<br />
can be checked and changed with the Dräger CC Vision software.<br />
First commissioning<br />
Quick bump test<br />
On<br />
On<br />
allowed<br />
4.4 (vol. %)<br />
(4.4 vol. % corresponds to 100 %LEL)<br />
15 minutes for STEL<br />
8 hours for TWA<br />
Before using the instrument for the first time, insert the batteries supplied or<br />
a charged T4 NiMH power pack T4 (HBT 0000 type, order no. 83 18 704) /<br />
T4 HC (HBT 0100 type, order no. 83 22 244) (refer to chapter "Replacing the<br />
batteries" on Page 15). The Dräger X-<strong>am</strong> <strong>5000</strong> is ready for operation.<br />
Operation<br />
Switching on the instrument<br />
• Press and hold the [OK] key for approx. 3 seconds until the countdown<br />
»3.2.1« shown on the display has elapsed.<br />
– All the display segments, including the visual, audible and vibration alarms,<br />
are activated for a short time.<br />
– The software version is displayed.<br />
– The instrument performs a self test.<br />
– The next sensor to be calibrated is displayed, along with the remaining days<br />
to the next calibration, e. g. » Ex%LEL CAL20«.<br />
– The time until the bump test interval elapses is displayed in days, e.g. » bt 123 «.<br />
12