10.4 Accuracy<br />

Accuracy of the method has been checked by parallelism test.<br />

Parallelism test<br />

A medium TGAb concentration serum was tested at different dilutions with the<br />

Diluent sample.<br />

Dilution Expected<br />

Measured<br />

(IU/mL)<br />

(IU/mL)<br />

S1 undiluted 225.99<br />

1: 2 112.99 86.1<br />

1: 4 56.49 54.42<br />

1: 8 28.24 20.72<br />

1: 16 14.12 12.91<br />

10.5 High-dose Effect<br />

High-dose effect was verified with samples containing high antibody<br />

concentrations .<br />

Concentrations of IgG to Tg ≤ 3.000.000 IU/ml were observed with B/T higher<br />

than last calibrator concentration .<br />

11. LIMITS OF THE ASSAY<br />

The results of the assay must be carefully interpreted and confirmed by clinical<br />

evaluations and further diagnostic tests.<br />

12. SYMBOLS LEGEND: see pag. 19<br />

KT8<strong>CT</strong> – <strong>TgAb</strong> irma <strong>CT</strong><br />

M91 – Rev.01 – 09/2004 – Pag. 18/24