APPENDIX 2 - DANETV
APPENDIX 2 - DANETV
APPENDIX 2 - DANETV
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
opløsningsmiddelsammensætningen i gassen med henblik på at beregne en resulterende<br />
responsfaktor og en resulterende omregningsfaktor.<br />
Ranges: 0 - 30, 0 - 300, 0 – 3.000, 0 – 30.000, 0 – 300.000 ppm C<br />
Limit of detection: 1 mg C/m³<br />
Uncertainty: 5 % of measured value (95% confidense interval).<br />
FORCE Technology method: EM-12-01<br />
Reference/standard: EN 12619, EN 13526, VDI 3481 bl.3<br />
ELPI measuring particle number and size distribution<br />
The ELPI (Electrical Low Pressure Impactor) sampling technique is a real time measurement of<br />
the number concentration and size distribution through 12 particle size intervals (from 7nm to<br />
10,000 nm). Particles are collected from the submicron range (from 30 nm) on 11 individual<br />
substrates according to their aerodynamic size. The ELPI makes a real-time measurement of<br />
each impactor stage by charging the particles before the impactor, and by counting the number<br />
of particles collected at each stage. The mass concentration can be calculated assuming a unit<br />
density, e.g. 1 g/cm³, for all particles.<br />
When measuring high concentrations of particles, a dilution can be necessary. Sampling can be<br />
performed through a dilution system comprising of a two-step dilution (dilution factor<br />
approximately 1:100). Dry, HEPA filtered pressurized air is added as dilution air to the diluters.<br />
The actual dilution ratio can be measured by means of propane gas and a flame ionization<br />
detector. The diluted gas will reached a temperature slightly above ambient before the inlet of<br />
the ELPI instrument.<br />
The effect of non-isokinetic sampling is rather small for particles < 2500 nm, however isokinetic<br />
sampling is sought. An out-stack pre-separator (cyclone) for separation of particles > 10 µm was<br />
used.