<strong>JAEA</strong>-<strong>Technology</strong> <strong>2010</strong>-005<br />

Establishment of Soft X-ray Reference Fields for Performance Tests of<br />

Radiation Measuring Instruments Based on National Standard<br />

Shigeru SHIMIZU, Katsutoshi FUJII and Tomokatsu KAWASAKI +<br />

Department of Radiation Protection<br />

Nuclear Science Research Institute<br />

Tokai Research and Development Center<br />

Japan Atomic Energy Agency<br />

Tokai-mura, Naka-gun, Ibaraki-ken<br />

(Received January 26 , <strong>2010</strong>)<br />

In the Japanese Industrial Standard, JIS Z 4511 sets the condition of X-ray reference fields to<br />

be used for the performance test of radiation measuring instruments for radiation protection.<br />

With the soft X-ray generator installed in the Facility of Radiation Standards in Nuclear Science<br />

Research Institute, we have established four series of soft X-ray reference fields with quality<br />

index of 0.6, 0.7, 0.8 and 0.9 based on JIS Z 4511. In this article, quality of the X-ray fields,<br />

X-ray spectra, conversion coefficients to the dose equivalents from Air-Kerma were evaluated in<br />

the soft X-ray fields produced with X-ray tube voltage ranging from 6kV to 100kV, and set soft<br />

X-ray reference fields based on Japanese Industrial Standard about 42 qualities. As a result,<br />

this X-ray reference fields met a domestic setting condition well and precision was good for<br />

conversion factors to various dose equivalents from Air- Kerma and we confirmed the soundness<br />

of the spectrum of each X-ray quality. By this, it is found that dose equivalent standard with<br />

good precision and wide range of test energy points and dose rates can be provided, for the<br />

performance tests such as the energy characteristic and the direction characteristic for various<br />

radiation instruments.<br />

Keywords: X-ray Reference Field, Radiation Measuring Instrument, JIS Z 4511,<br />

Half Value Layer, Quality Index, X-ray Energy Spectra, Calibration,<br />

Effective Energy, National Standard,<br />