<strong>JAEA</strong>-<strong>Technology</strong> <strong>2010</strong>-005<br />
Establishment of Soft X-ray Reference Fields for Performance Tests of<br />
Radiation Measuring Instruments Based on National Standard<br />
Shigeru SHIMIZU, Katsutoshi FUJII and Tomokatsu KAWASAKI +<br />
Department of Radiation Protection<br />
Nuclear Science Research Institute<br />
Tokai Research and Development Center<br />
Japan Atomic Energy Agency<br />
Tokai-mura, Naka-gun, Ibaraki-ken<br />
(Received January 26 , <strong>2010</strong>)<br />
In the Japanese Industrial Standard, JIS Z 4511 sets the condition of X-ray reference fields to<br />
be used for the performance test of radiation measuring instruments for radiation protection.<br />
With the soft X-ray generator installed in the Facility of Radiation Standards in Nuclear Science<br />
Research Institute, we have established four series of soft X-ray reference fields with quality<br />
index of 0.6, 0.7, 0.8 and 0.9 based on JIS Z 4511. In this article, quality of the X-ray fields,<br />
X-ray spectra, conversion coefficients to the dose equivalents from Air-Kerma were evaluated in<br />
the soft X-ray fields produced with X-ray tube voltage ranging from 6kV to 100kV, and set soft<br />
X-ray reference fields based on Japanese Industrial Standard about 42 qualities. As a result,<br />
this X-ray reference fields met a domestic setting condition well and precision was good for<br />
conversion factors to various dose equivalents from Air- Kerma and we confirmed the soundness<br />
of the spectrum of each X-ray quality. By this, it is found that dose equivalent standard with<br />
good precision and wide range of test energy points and dose rates can be provided, for the<br />
performance tests such as the energy characteristic and the direction characteristic for various<br />
radiation instruments.<br />
Keywords: X-ray Reference Field, Radiation Measuring Instrument, JIS Z 4511,<br />
Half Value Layer, Quality Index, X-ray Energy Spectra, Calibration,<br />
Effective Energy, National Standard,<br />
+:Health and Safety Department, Oarai Research and Development Center<br />
ii