4 Installation<br />

4.6 Connecting the Unit<br />

<br />

• Insert the flexible single branch liquid light guide with the Ø16mm fitting into the light<br />

guide socket 1 in the front panel of the light source until it seats with an audible click.<br />

Shutter will not operate unless light guide is fully inserted.<br />

When using a dual branch liquid light<br />

guide the two branches must be<br />

positioned vertically above each other in<br />

the light guide socket in the front panel.<br />

• Insert the light guide all the way into the<br />

socket 1 and rotate it until it snaps into<br />

vertical position.<br />

<strong>97034</strong><br />

Be sure that no fingerprints are neither on the light entry nor the light exit ends of the<br />

light guide.<br />

Fingerprints can be removed with a lint free wipe and pure alcohol.<br />

Any contamination, which has not been removed, will bake to the light guide and result<br />

in premature deterioration of performance.<br />

• Connect power cord and plug the unit into a properly grounded AC outlet.<br />

• Connect footswitch on the backside of the system, if it is required.