Lamp Life Time<br />
Based on an average power-on time of 8 hours per day, the lamp has an expected<br />
operating lifetime of 1500 hours. UV lamp intensity deteriorates gradually with time.<br />
Lamp life depends very much on length of operating intervals, as each ignition process<br />
wears the lamp considerably.<br />
Frequent on/off switching results in premature deterioration of the lamp. Do not<br />
switch off for idle periods of less than 3 hours.<br />
The lamp contains mercury. Do not discard with regular trash.<br />
Disposal as hazardous waste required.<br />
To ensure uniform curing process control, output of UV radiation from the system is<br />
monitored with an integrated UV meter. A lower UV threshold setting can be defined to<br />
prevent imperfectly cured bonds. A setting should be selected which will achieve the<br />
desired curing results and still prolong the lifetime of the lamp.<br />
Lamp Handling<br />
Any contamination, such as fingerprints on the glass, which has not been removed will<br />
bake to the envelope and result in premature deterioration of performance.<br />
Fingerprints can be removed with a lint free wipe and pure alcohol.<br />
Sometimes the envelope can burst in rare occasions – especially if the lamp is very old.<br />
The reasons could be etched fingerprints on the envelope or too frequent on/off<br />
switching.<br />
Broken lamps will release mercury.<br />
Strategy:<br />
Remove every person close by the UV Wand System at once to prevent them from<br />
inhaling of mercury vapors. The room should be ventilated well (15 – 20 minutes).<br />
After cooling off of the system the residues of the mercury should be removed<br />
with absorption tools, which can be bought from specialized trades of chemistry.<br />
Overheating<br />
If the maximum operating temperature is exceeded the lamp is automatically switched<br />
off and the red LED 5 lights up, and an error message appears on the LCD display. In<br />
addition, an acoustic signal will be emitted. The fan continues to run. After the unit has<br />
cooled off sufficiently the error message will extinguish and the lamp can be ignited<br />
again. Reasons for overheating can be obstruction of ventilation slots or high ambient<br />
temperatures. The rotation of the fan is also monitored electronically.