1 Introduction<br />

1.1 Intended use<br />

The Précise 1100 SD is a mains-operated measuring unit which can also be<br />

used with batteries. It is used for determining oxygen content in percentage<br />

form, for flow measurement at the output of oxygen concentrators and for<br />

pressure measurement of non-aggressive gases (various pressure<br />

measurement range options available). The measurement log and the<br />

graphical displays can be printed out using the thermal printer.<br />

Important notice!<br />

The Précise 1100 SD may not be used for monitoring life support systems.<br />

The oxygen concentration may only be measured in an oxygen + nitrogen<br />

gas mix (e.g. oxygen concentrator).<br />

1.2 Function description<br />

An ultrasonic sensor is used to measure the oxygen concentration and flow.<br />

This sensor needs no recalibration, although Medicap homecare<br />

recommends that the unit measurements be checked annually.<br />

The oxygen concentration measurement is based on the different sound<br />

propagation times in oxygen and nitrogen. Flow measurement is carried out<br />

by analysing different sound propagation times in and against the direction<br />

of flow.<br />

Pressure is measured by a pressure cell.<br />

1.3 Important instructions for safety and use<br />

The precondition for correct use of the Précise 1100 SD measuring unit is<br />

thorough knowledge of and compliance with these instructions for use.<br />

Operating manual P1100SD Rev. D 2012-01 3