1 Introduction<br />
1.1 Intended use<br />
The Précise 1100 SD is a mains-operated measuring unit which can also be<br />
used with batteries. It is used for determining oxygen content in percentage<br />
form, for flow measurement at the output of oxygen concentrators and for<br />
pressure measurement of non-aggressive gases (various pressure<br />
measurement range options available). The measurement log and the<br />
graphical displays can be printed out using the thermal printer.<br />
Important notice!<br />
The Précise 1100 SD may not be used for monitoring life support systems.<br />
The oxygen concentration may only be measured in an oxygen + nitrogen<br />
gas mix (e.g. oxygen concentrator).<br />
1.2 Function description<br />
An ultrasonic sensor is used to measure the oxygen concentration and flow.<br />
This sensor needs no recalibration, although Medicap homecare<br />
recommends that the unit measurements be checked annually.<br />
The oxygen concentration measurement is based on the different sound<br />
propagation times in oxygen and nitrogen. Flow measurement is carried out<br />
by analysing different sound propagation times in and against the direction<br />
of flow.<br />
Pressure is measured by a pressure cell.<br />
1.3 Important instructions for safety and use<br />
The precondition for correct use of the Précise 1100 SD measuring unit is<br />
thorough knowledge of and compliance with these instructions for use.<br />
Operating manual P1100SD Rev. D 2012-01 3