2.3 Switching on the unit<br />
printer<br />
shuttle knob<br />
Fig. 3: Précise 1100 SD from the front<br />
To switch on, press the shuttle knob and hold (at least 3 sec.). After<br />
switching on, the following backlit display appears giving the current<br />
oxygen concentration in figures.<br />
The graphical display can be displayed via plotter (see section 2.3 Menu<br />
settings). If the oxygen concentration is less than 70% by volume, the<br />
display shows "--.- %/vol".<br />
Please note that most oxygen concentrators do not give precise<br />
concentration measurements until at least 10 min. after being switched on.<br />
Do not connect a humidifier between the oxygen concentrator and the<br />
Précise 1100 SD.<br />
Operating manual P1100SD Rev. D 2012-01 7