Supplement 45 Ultrasound Protocol Support - Dicom - NEMA
Supplement 45 Ultrasound Protocol Support - Dicom - NEMA
Supplement 45 Ultrasound Protocol Support - Dicom - NEMA
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>Supplement</strong> <strong>45</strong>: <strong>Ultrasound</strong> <strong>Protocol</strong><br />
Page 5<br />
Patient Patient C.7.1.1 M<br />
Study General Study C.7.2.1 M<br />
Patient Study C.7.2.2 U<br />
Series General Series C.7.3.1 M<br />
Frame of<br />
Reference<br />
US Series C.8.5.x U – Suggested if image is<br />
part of an acquisition<br />
protocol<br />
Frame of Reference C.7.4.1 U<br />
US Frame of Reference C.8.5.4 C - Required if images are<br />
spatially related<br />
Equipment General Equipment C.7.5.1 M<br />
Image General Image C.7.6.1 M<br />
(See A.6.4.1) Image Pixel C.7.6.3 M<br />
Contrast/bolus C.7.6.4 C - Required if contrast<br />
media was used in this<br />
image<br />
Palette Color Lookup table C.7.9 C - Required if Photometic<br />
Interpretation (0028,0004)<br />
has a value of PALETTE<br />
COLOR<br />
US Region Calibration C.8.5.5 U<br />
US Image C.8.5.6 M<br />
US Staged <strong>Protocol</strong> C.8.5.y U – Suggested if image is<br />
part of a staged protocol<br />
Overplay Plane C.9.2 U<br />
VOI LUT C.11.2 U<br />
SOP Common C.12.1 M<br />
Curve Curve Identification C.10.1 M<br />
(See A.6.4.1) Curve C.10.2 M<br />
Audio C.10.3 U<br />
SOP Common C.12.1 M<br />
A.6.4.1 Mutually Exclusive IEs<br />
The Image and Curve IEs are mutually exclusive. Each SOP Instance using this IOD shall contain exactly<br />
one of these IEs.<br />
A.6.5 US Image IOD Relationship to the DICOM Real-World Model<br />
<strong>Ultrasound</strong> images that make use of the US Series & <strong>Protocol</strong> Modules have a more specific<br />
relationship to the DICOM Real-World Model than non-protocol images. An ultrasound acquisition<br />
protocol is considered one Requested Procedure and has one Procedure Plan. The Procedure<br />
Plan will have one or more Action Items. An ultrasound acquisition protocol will typically consist