C.8.5.y.1.3 Transducer Orientation Sequence<br />
<strong>Supplement</strong> <strong>45</strong>: <strong>Ultrasound</strong> <strong>Protocol</strong><br />
Page 13<br />
Transducer Orientation Sequence (0008,2244) identifies the transducer orientation used in this image.<br />
The Coding Scheme Designator (0008,0102) shall be SNM3.<br />
The Code Value (0008,0100) shall be drawn from the SNOMED DICOM Microglossary Context ID 6.<br />
Transducer Orientation Modifier Sequence (0008,2246) modifies the transducer orientation used in this<br />
image.<br />
The Coding Scheme Designator (0008,0102) shall be SNM3.<br />
The Code Value (0008,0100) shall be drawn from the SNOMED DICOM Microglossary Context ID 7.<br />
C.8.5.y.1.4 Stages<br />
The Number of Stages (0008,2124) shall be the total number of stages performed in this instance of the<br />
protocol. Stage Number (0008,2122) shall be the number that indicates the default order in which the<br />
stages are to be presented. Stage Number shall start from 1.<br />
C.8.5.y.1.5 Views<br />
The View Number (0008,2128) shall be used identify the same view across stages and defines a default<br />
order for presentation within and across stages. If a particular view appears in one stage but not in a<br />
subsequent stage, then that View Number will not be used during that subsequent stage. View Number<br />
shall start from 1. The Number of Views in Stage (0008,212A) shall be used to indicate how many unique<br />
View Numbers exist for this stage.<br />
The addition of modification of the value for this attribute shall not by itself force a SOP Instance UID<br />
change.