of one Scheduled Procedure Step containing one or more Action Items. Assuming the acquisition<br />
protocol was successfully completed, this will result in one Modality Performed Procedure Step.<br />
Therefore, all images that are part of this particular instance of an ultrasound acquisition protocol<br />
will have the same Procedure Code Sequence (0032,1064). It is recommended that all the images<br />
that are part of this particular instance of an ultrasound acquisition protocol have the same<br />
Performed Procedure Step ID. In an environment with information system connectivity, the<br />
Procedure Code Sequence can be used to identify all the images that were acquired as part of that<br />
protocol. If no information system connectivity is available, then Performed Procedure Step ID<br />
may be used to identify all the images that were acquired as part of that protocol.<br />
It is recommended that each stage should map to a separate series.<br />
Item: Add to PS 3.3 Table A.7.4-1 - for clarity, those modules that are changed from the existing standard<br />
or new are shown in Bold and Underlined<br />
A.7.4 US Multi-Frame Image IOD Module Table<br />
Table A.7.4-1<br />
US MULTI-FRAME IMAGE IOD MODULES<br />
IE Module Reference Usage<br />
Patient Patient C.7.1.1 M<br />
Study General Study C.7.2.1 M<br />
Patient Study C.7.2.2 U<br />
Series General Series C.7.3.1 M<br />
Frame of<br />
Reference<br />
US Series C.8.5.x U – Suggested if image is<br />
part of an acquisition<br />
protocol<br />
Frame of Reference C.7.4.1 U<br />
US Frame of Reference C.8.5.4 C - Required if images are<br />
spatially related<br />
Equipment General Equipment C.7.5.1 M<br />
Image General Image C.7.6.1 M<br />
(See A.7.4.1) Image Pixel C.7.6.3 M<br />
Contrast/bolus C.7.6.4 C - Required if contrast<br />
media was used in this<br />
image.<br />
Cine C.7.6.5 M<br />
Multi-frame C.7.6.6 M<br />
Palette Color Lookup Table C.7.9 C - Required if Photometric<br />
Interpretation (0028,0004)<br />
has a value of PALETTE<br />
COLOR<br />
6