24.07.2013 Views

Supplement 45 Ultrasound Protocol Support - Dicom - NEMA

Supplement 45 Ultrasound Protocol Support - Dicom - NEMA

Supplement 45 Ultrasound Protocol Support - Dicom - NEMA

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

of one Scheduled Procedure Step containing one or more Action Items. Assuming the acquisition<br />

protocol was successfully completed, this will result in one Modality Performed Procedure Step.<br />

Therefore, all images that are part of this particular instance of an ultrasound acquisition protocol<br />

will have the same Procedure Code Sequence (0032,1064). It is recommended that all the images<br />

that are part of this particular instance of an ultrasound acquisition protocol have the same<br />

Performed Procedure Step ID. In an environment with information system connectivity, the<br />

Procedure Code Sequence can be used to identify all the images that were acquired as part of that<br />

protocol. If no information system connectivity is available, then Performed Procedure Step ID<br />

may be used to identify all the images that were acquired as part of that protocol.<br />

It is recommended that each stage should map to a separate series.<br />

Item: Add to PS 3.3 Table A.7.4-1 - for clarity, those modules that are changed from the existing standard<br />

or new are shown in Bold and Underlined<br />

A.7.4 US Multi-Frame Image IOD Module Table<br />

Table A.7.4-1<br />

US MULTI-FRAME IMAGE IOD MODULES<br />

IE Module Reference Usage<br />

Patient Patient C.7.1.1 M<br />

Study General Study C.7.2.1 M<br />

Patient Study C.7.2.2 U<br />

Series General Series C.7.3.1 M<br />

Frame of<br />

Reference<br />

US Series C.8.5.x U – Suggested if image is<br />

part of an acquisition<br />

protocol<br />

Frame of Reference C.7.4.1 U<br />

US Frame of Reference C.8.5.4 C - Required if images are<br />

spatially related<br />

Equipment General Equipment C.7.5.1 M<br />

Image General Image C.7.6.1 M<br />

(See A.7.4.1) Image Pixel C.7.6.3 M<br />

Contrast/bolus C.7.6.4 C - Required if contrast<br />

media was used in this<br />

image.<br />

Cine C.7.6.5 M<br />

Multi-frame C.7.6.6 M<br />

Palette Color Lookup Table C.7.9 C - Required if Photometric<br />

Interpretation (0028,0004)<br />

has a value of PALETTE<br />

COLOR<br />

6

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!