Option A: ABLB—Alternate Binaural Loudness Balance<br />

This is a test for binaural recruitment. Recruitment is an abnormally large increase in hearing<br />

sensation for a given increase in sound level. Derecruitment is an abnormally small increase<br />

in hearing sensation for a given increase in sound level.<br />

This test will automatically alternate a tone of a selectable frequency between the left ear and<br />

right ear. The tone is presented for:<br />

appoximately 370 milliseconds in the left ear,<br />

approximately 130 milliseconds of silence,<br />

approximately 370 milliseconds in the right ear,<br />

approximately 130 milliseconds of silence<br />

(the pattern repeats).<br />

ABLB Setup:<br />

right <strong>Hearing</strong> Aid Simulator: Option A<br />

left <strong>Hearing</strong> Aid Simulator: Off<br />

right Output: Phone<br />

left Output: Phone<br />

left Input: Stenger<br />

right Input: Tone<br />

Pulse: On<br />

left simulus Reverse: On<br />

right Stimulus: Off (does not matter)<br />

frequency: (select desired frequency)<br />

left <strong>Hearing</strong> Level: (select desired amplitude)<br />

right <strong>Hearing</strong> Level: (select desired amplitude)<br />

Perform the desired ABLB hearing test.<br />

Notes:<br />

Many other Option A, ABLB setups are possible. You may select different inputs and outputs.<br />

Option A, ABLB works by disabling the Stimulus for one channel for 370 milliseconds, plus<br />

130 milliseconds of silence (both channels), then disabling the other channel for 370 milliseconds<br />

plus 130 milliseconds of silence (both channels).<br />

Option A, ABLB is active when all four of these conditions are met:<br />

• Left <strong>Hearing</strong> Aid Simulator switch set to Off<br />

• Right <strong>Hearing</strong> Aid Simulator switch set to Option A ˛<br />

• Pulse is ON<br />

• Stimulus is enabled for both channels or stimulus is enabled for one channel if Stenger<br />

is selected.<br />

If Pulse is turned off, the <strong>FA</strong>-<strong>10</strong> will operate as if the right <strong>Hearing</strong> Aid Simulator switch is<br />

set to Off.<br />