DICOM Conformance Statement Page 18 of 94<br />
Source Reason/Diagnosis Behavior<br />
4 – unrecognized-PDU<br />
parameter<br />
5 – unexpected-PDU<br />
parameter<br />
6 – invalid-PDUparameter<br />
value<br />
When received, the Eleva Workspot<br />
terminates the connection with the following<br />
log: Association ABORTED by peer (<br />
2: ABORT_SOURCE_dul_provider,<br />
4: ABORT_REASON<br />
_unrecognized_pdu_parameter).<br />
When received, the Eleva Workspot<br />
terminates the connection with the following<br />
log: Association ABORTED by peer (<br />
2: ABORT_SOURCE_dul_provider,<br />
5: ABORT_REASON<br />
_unexpected_pdu_parameter).<br />
When received, the Eleva Workspot<br />
terminates the connection with the following<br />
log: Association ABORTED by peer (<br />
2: ABORT_SOURCE_dul_provider,<br />
6: ABORT_REASON<br />
_invalid_pdu_parameter).<br />
The behavior of the AE during DICOM communication failure is summarized in next<br />
table.<br />
Table 13: DICOM Command Communication Failure Behavior<br />
Exception Behavior<br />
Timeout The Association is aborted using A-ABORT and command marked as failed.<br />
The reason is logged and reported to the user.<br />
Association aborted The command is marked as failed. The reason is logged and reported to the<br />
user.<br />
© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />
Document Number: XPB 031- 09006<strong>1.0</strong>2 <strong>Essenta</strong> <strong>DR</strong> <strong>Compact</strong> <strong>1.0</strong> 2 Apr 2009