02.08.2013 Views

Essenta DR Compact 1.0 - InCenter - Philips

Essenta DR Compact 1.0 - InCenter - Philips

Essenta DR Compact 1.0 - InCenter - Philips

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

DICOM Conformance Statement Page 18 of 94<br />

Source Reason/Diagnosis Behavior<br />

4 – unrecognized-PDU<br />

parameter<br />

5 – unexpected-PDU<br />

parameter<br />

6 – invalid-PDUparameter<br />

value<br />

When received, the Eleva Workspot<br />

terminates the connection with the following<br />

log: Association ABORTED by peer (<br />

2: ABORT_SOURCE_dul_provider,<br />

4: ABORT_REASON<br />

_unrecognized_pdu_parameter).<br />

When received, the Eleva Workspot<br />

terminates the connection with the following<br />

log: Association ABORTED by peer (<br />

2: ABORT_SOURCE_dul_provider,<br />

5: ABORT_REASON<br />

_unexpected_pdu_parameter).<br />

When received, the Eleva Workspot<br />

terminates the connection with the following<br />

log: Association ABORTED by peer (<br />

2: ABORT_SOURCE_dul_provider,<br />

6: ABORT_REASON<br />

_invalid_pdu_parameter).<br />

The behavior of the AE during DICOM communication failure is summarized in next<br />

table.<br />

Table 13: DICOM Command Communication Failure Behavior<br />

Exception Behavior<br />

Timeout The Association is aborted using A-ABORT and command marked as failed.<br />

The reason is logged and reported to the user.<br />

Association aborted The command is marked as failed. The reason is logged and reported to the<br />

user.<br />

© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />

Document Number: XPB 031- 09006<strong>1.0</strong>2 <strong>Essenta</strong> <strong>DR</strong> <strong>Compact</strong> <strong>1.0</strong> 2 Apr 2009

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!