DICOM Conformance Statement Page 38 of 94<br />

4.2.1.3.4.3.4. Dataset Specific Conformance for Modality Worklist<br />

(Broadcast Query) C-FIND SCU<br />

This section includes the dataset specific behavior, i.e. error codes, error and<br />

exception handling, time-outs, etc.<br />

The table below shall be read as follows:<br />

Attribute Name: Attributes supported to build a Modality Worklist Request Identifier.<br />

Tag: DICOM tag for this attribute.<br />

VR: DICOM VR for this attribute.<br />

M: Matching Keys for (automatic) Worklist Update.<br />

R: Return Keys. An “X” will indicate that this attribute as Return Key<br />

with zero length for Universal Matching.<br />

Q: Interactive Query Key. An “X” will indicate that this attribute as<br />

matching key can be used.<br />

D: Displayed Keys. An “X” indicates that this Worklist attribute is<br />

displayed to the user during a patient registration dialog.<br />

IOD: An “X” indicates that this Worklist attribute is included into all object<br />

Instances created during performance of the related Procedure<br />

Step.<br />

Type of matching: The following types of matching exists:<br />

Single Value Matching<br />

List of UID Matching<br />

Wild Card Matching<br />

Range Matching<br />

Sequence Matching<br />

Universal Matching<br />

Table 35: Worklist Request Identifier<br />

Attribute Name Tag VR M R Q D IOD Type of Matching Comment<br />

SOP Common Module<br />

Specific Character Set 0008,0005 CS X X If Configured<br />

Patient Identification Module<br />

Patient's Name 0010,0010 PN X Last name, First name,<br />

Prefix, Middle name,<br />

Suffix<br />

Patient ID 0010,0020 LO X<br />

Issuer of Patient ID 0010,0021 LO<br />

Other Patient IDs 0010,1000 LO<br />

Patient Demographic Module<br />

Patient's Birth Date 0010,0030 DA X<br />

Patient's Sex 0010,0040 CS X<br />

Patient's Age 0010,1010 AS<br />

Patient's Size 0010,1020 DS<br />

Patient's Weight 0010,1030 DS X<br />

Ethnic Group 0010,2160 SH<br />

Occupation 0010,2180 SH<br />

Patient Comments 0010,4000 LT<br />

Confidentiality Constraint on<br />

Patient Data Description<br />

0040,3001 LO<br />

Patient Medical Module<br />

Medical Alerts 0010,2000 LO X<br />

Contrast Allergies 0010,2110 LO<br />

Additional Patient History 0010,21B0 LT<br />

Pregnancy Status 0010,21C0 US X<br />

Special Needs 0038,0050 LO<br />

Current Patient Location 0038,0300<br />

Visit Status Module<br />

LO<br />

© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />

Document Number: XPB 031- 09006<strong>1.0</strong>2 <strong>Essenta</strong> <strong>DR</strong> <strong>Compact</strong> <strong>1.0</strong> 2 Apr 2009