Essenta DR Compact 1.0 - InCenter - Philips
Essenta DR Compact 1.0 - InCenter - Philips
Essenta DR Compact 1.0 - InCenter - Philips
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
DICOM Conformance Statement Page 38 of 94<br />
4.2.1.3.4.3.4. Dataset Specific Conformance for Modality Worklist<br />
(Broadcast Query) C-FIND SCU<br />
This section includes the dataset specific behavior, i.e. error codes, error and<br />
exception handling, time-outs, etc.<br />
The table below shall be read as follows:<br />
Attribute Name: Attributes supported to build a Modality Worklist Request Identifier.<br />
Tag: DICOM tag for this attribute.<br />
VR: DICOM VR for this attribute.<br />
M: Matching Keys for (automatic) Worklist Update.<br />
R: Return Keys. An “X” will indicate that this attribute as Return Key<br />
with zero length for Universal Matching.<br />
Q: Interactive Query Key. An “X” will indicate that this attribute as<br />
matching key can be used.<br />
D: Displayed Keys. An “X” indicates that this Worklist attribute is<br />
displayed to the user during a patient registration dialog.<br />
IOD: An “X” indicates that this Worklist attribute is included into all object<br />
Instances created during performance of the related Procedure<br />
Step.<br />
Type of matching: The following types of matching exists:<br />
Single Value Matching<br />
List of UID Matching<br />
Wild Card Matching<br />
Range Matching<br />
Sequence Matching<br />
Universal Matching<br />
Table 35: Worklist Request Identifier<br />
Attribute Name Tag VR M R Q D IOD Type of Matching Comment<br />
SOP Common Module<br />
Specific Character Set 0008,0005 CS X X If Configured<br />
Patient Identification Module<br />
Patient's Name 0010,0010 PN X Last name, First name,<br />
Prefix, Middle name,<br />
Suffix<br />
Patient ID 0010,0020 LO X<br />
Issuer of Patient ID 0010,0021 LO<br />
Other Patient IDs 0010,1000 LO<br />
Patient Demographic Module<br />
Patient's Birth Date 0010,0030 DA X<br />
Patient's Sex 0010,0040 CS X<br />
Patient's Age 0010,1010 AS<br />
Patient's Size 0010,1020 DS<br />
Patient's Weight 0010,1030 DS X<br />
Ethnic Group 0010,2160 SH<br />
Occupation 0010,2180 SH<br />
Patient Comments 0010,4000 LT<br />
Confidentiality Constraint on<br />
Patient Data Description<br />
0040,3001 LO<br />
Patient Medical Module<br />
Medical Alerts 0010,2000 LO X<br />
Contrast Allergies 0010,2110 LO<br />
Additional Patient History 0010,21B0 LT<br />
Pregnancy Status 0010,21C0 US X<br />
Special Needs 0038,0050 LO<br />
Current Patient Location 0038,0300<br />
Visit Status Module<br />
LO<br />
© Koninklijke <strong>Philips</strong> Electronics N.V. 2009<br />
Document Number: XPB 031- 09006<strong>1.0</strong>2 <strong>Essenta</strong> <strong>DR</strong> <strong>Compact</strong> <strong>1.0</strong> 2 Apr 2009