<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 28 of 78<br />
Figure 5: (Real World) Activity - Modality Performed Procedure Step As SCU<br />
Description of Activities<br />
A MammoDiagnost DR 1.0 "Examination" is regarded equivalent to a <strong>DICOM</strong><br />
Procedure Step. It is scheduled or manually entered before an acquisition is taken,<br />
and performed by taking acquisitions. If scheduled by the RIS, one Examination is the<br />
result of one Scheduled Procedure Step. Since an examination may not be re-opened<br />
after having been closed, and each examination workflow context is enclosed in one<br />
MPPS, one examination may result in 0:1 MPPS instances. However, image archiving<br />
after the examination's closure leads to 1:n MPPS instances per examination (append<br />
case).<br />
After the image for a Scheduled Procedure Step has been acquired, the system sets<br />
the MPPS status of the related examination to "IN PROGRESS" and generates an<br />
initial MPPS IN PROGRESS message. The system does not generate intermediate<br />
MPPS IN PROGRESS messages for subsequent acquisitions of this Scheduled<br />
Procedure Step instance.<br />
After finishing the appropriate acquisition(s), the system will change the MPPS status<br />
of the related examination to "COMPLETED" and generate an MPPS N-SET-FINAL<br />
message.<br />
MammoDiagnost DR 1.0 also generates MPPS messages for unscheduled<br />
examinations.<br />
The MPPS COMPLETED message will list the UID's of all related <strong>DICOM</strong> archived<br />
images and the format of (optionally) generated direct prints.<br />
After abandoning or discontinuing a procedure step, the operator may set the MPPS<br />
status of the related examination to "DISCONTINUED" and the system generates a<br />
© Koninklijke <strong>Philips</strong> Electronics N.V.<br />
Document Number: XPR 031-090167.00 MammoDiagnost DR 1.0 18-March-2009