<strong>DICOM</strong> <strong>Conformance</strong> <strong>Statement</strong> Page 30 of 78<br />

ELEVA AE by default derives the specific acquisition protocol from the Scheduled<br />

Protocol Code Sequence Items. Any single Item results in an Examination.<br />

ELEVA AE supports 3 more (configurable) mapping relations, as shown below:<br />

• Examination is selected from Scheduled Protocol Code Items->Code Value<br />

(0040,0008) (default)<br />

• Examination is selected from Scheduled Procedure Step Description<br />

(0040,0007)<br />

• Examination is selected from Requested Procedure Code Items->Code Value<br />

(0032,1064)<br />

• Examination is selected from Requested Procedure Description (0032,1060)<br />

ELEVA AE does not evaluate the attributes Coding Scheme Version (0008,0103),<br />

Coding Scheme Designator (0008,0102), Code Meaning (0008,0104), but only the<br />

Code Value (0008,0100), for mapping the examination settings. I.e. ELEVA AE<br />

expects that any used Code Value is unique (unambiguous) within a given RIS<br />

domain.<br />

Restrictions Depending on Number of Scheduled Protocol Code Items<br />

It is highly recommended that the Scheduled Procedure Step contains only 1 Item in<br />

the Scheduled Protocol Code Sequence.<br />

If the Scheduled Procedure Step contains items in the Scheduled Protocol Code<br />

Sequence, the Scheduled Procedure Step is split into examinations, where any<br />

single examination shows only 1 of the Scheduled Protocol Code Items, but all the<br />

other attributes are the same.<br />

When such an examination is returned back via MPPS, also the Performed Protocol<br />

Code Sequence will show only 1 item. If all Scheduled Procedure Step Code<br />

Items are performed, MPPS instances will be sent back to the RIS, and the sum<br />

of all Performed Protocol Code Items will be .<br />

4.2.1.3.3.3.1. Dataset Specific <strong>Conformance</strong> for Modality Performed<br />

Procedure Step N-CREATE SCU<br />

This section includes the dataset specific behavior, i.e. error codes, error and<br />

exception handling, time-outs, etc.<br />

Table 25: MPPS Request Identifiers for N-CREATE-RQ<br />

Attribute Name Tag VR Value Comment<br />

Performed Procedure Step Information Module<br />

Performed Location 0040,0243 SH EMPTY<br />

Performed Procedure Step<br />

Description<br />

0040,0254 LO<br />

Performed Procedure Step End<br />

Date<br />

0040,0250 DA Finish of the examination<br />

Performed Procedure Step End<br />

Time<br />

0040,0251 TM Finish of the examination<br />

Performed Procedure Step ID 0040,0253 SH<br />

Performed Procedure Step Start<br />

Date<br />

0040,0244 DA Start of the examination<br />

Performed Procedure Step Start<br />

Time<br />

0040,0245 TM Start of the examination<br />

Performed Procedure Step<br />

Status<br />

0040,0252 CS<br />

Performed Procedure Type<br />

Description<br />

0040,0255 LO<br />

Performed Station AE Title 0040,0241 AE<br />

© Koninklijke <strong>Philips</strong> Electronics N.V.<br />

Document Number: XPR 031-090167.00 MammoDiagnost DR 1.0 18-March-2009