HIn3637 <strong>Manual</strong> Page — 15<br />

time. The HIn3637 should be charged only with an<br />

approved battery charger.<br />

Battery Replacement<br />

A NiCd battery powers the HIn3637 Magnetic Field<br />

Meter (See Table 5n1). In the event of battery failure, a<br />

new battery can be obtained from Holaday Industries.<br />

NOTE:<br />

A #0 Phillips screwdriver works well for the<br />

following procedure.<br />

To replace the battery, remove the eight screws located<br />

around the lip of the meter's front panel, pull the panel<br />

away from the chassis and unplug the battery connector.<br />

Remove the battery retainer and the old battery. Install<br />

the new battery and reverse the above steps to<br />

reassemble the meter.<br />

CAUTION<br />

Never short the terminals of a NiCd battery (even a<br />

discharged battery). The resulting current may be excessive<br />

and cause the lead wires to become extremely hot, which<br />

can result in severe burns and/or fire.