HIn3637 <strong>Manual</strong> Page — 27<br />

8.0 CALIBRATION<br />

The HIn3637 is calibrated using a Helmholtz coil<br />

(consisting of two close-coupled windings) one meter in<br />

diameter to establish an accurate, known magnetic field<br />

flux density. A precisely controlled and measured<br />

sinusoidal current is driven through the coils and, based<br />

on the dimensions of the coils, the magnetic field flux<br />

density between the coils is calculated.<br />

Each axis is individually calibrated. Calibration is<br />

performed at a frequency of 20 kHz, and verified at 232<br />

kHz, at a flux density level of 10 mG. The out of band<br />

response is checked at 60 Hz and 700 kHz. All calibration<br />

is traceable to NIST.<br />

The HIn3637 indicates magnetic flux density (B) in units<br />

of Tesla/Gauss. The flux density in microteslas or the<br />

magnetic field strength (H) in milliamperes per meter are<br />

related as follows:<br />

1 mG = 0.1 microtesla (:T)<br />

1 mG = 80 milliamperes per meter (mA/m)<br />

The recommended calibration interval is one year.<br />

Holaday Industries offers complete calibration services<br />

which comply with MILnSTD 45662A.