Sample Environment<br />

• Rheometer shear sandwich<br />

• Rheowis-fluid rheometer (max. shear rate<br />

10000 s -1 )<br />

• Anton-Paar fluid rheometer<br />

• Stopped flow cell<br />

• Sample holders: 9 horizontal x 3 vertical (temperature<br />

controlled) for standard Hellma cells<br />

404.000-QX, and 110-QX<br />

• Oil & water thermostats (typical 10..100°C)<br />

• Electric thermostat (RT..200°C)<br />

• 6-positions thermostated (Peltier) sample<br />

holder (-40°C..150°C)<br />

• Magnet (horizontal, vertical)<br />

• Cryostat with sapphire windows<br />

• High temperature furnace<br />

• Pressure cells (500 bar, 2000 bar, 5000 bar)<br />

Technical Data<br />

Overall performance<br />

Q = 0.0007..0.5 Å-1 Maximal flux: 3 x 108 n/cm2 /s<br />

Typical flux: 8 x 106 n/cm2 /s<br />

(collimation 8 m, aperture 30 x 30 mm², λ = 7 Å)<br />

Velocity selector<br />

Dornier, FWHM 20 %, λ = 4.5 Å – 12 Å, 20 Å<br />

Chopper<br />

for TOF-wavelength analysis, FWHM 1 %<br />

Active apertures<br />

2 m, 4 m, 8 m, 14 m, 20 m<br />

Aperture sizes<br />

rectangular 1 x 1 mm2 – 50 x 50 mm2 Sample aperture<br />

rectangular 1 x 1 mm2 – 50 x 50 mm2 Dr. Aurel Radulescu<br />

Phone: +49.(0)89.289.10712<br />

Email: a.radulescu@fz-juelich.de<br />

Phone Instrument: .14326<br />

Figure 3: Time-resolved SANS (1s time resolution) of fast chain<br />

exchange under equilibrium between PEP(1k)-PEO(1k) cylindrical<br />

micelles using a stopped-flow apparatus at KWS-2<br />

<strong>Neutron</strong> lenses<br />

MgF , diameter 50mm, curvature 20mm<br />

2<br />

Packs with 4, 6,16 lenses<br />

Sample stage<br />

XYZθ translational-rotational stage<br />

Accuracy better than 0.01°, 0.01mm<br />

Detector 1<br />

Detection range: continuous 1 m – 20 m<br />

6Li-Scintillator 1 mm thickness + Photomultiplier<br />

Efficiency better than 95 %, Spatial resolution<br />

5.3 x 5.3 mm2 , 128 x 128 channels<br />

Max. countrate 0.6 MHz (τ = 0.64 μs)<br />

dead<br />

Detector 2 (high res.)<br />

Spatial resolution 0.45 x 0.45 mm2 Active area: Ø = 8.7 cm<br />

6Li-Scintillator 1 mm thickness<br />

Fixed position: 17 m after sample position<br />

Dr. Vitaly Pipich<br />

Phone: +49.(0)89.289.10710<br />

Email: v.pipich@fz-juelich.de<br />

www.jcns.info/jcns_kws2<br />

SANS<br />

34 / 35