Sample Environment<br />
• Rheometer shear sandwich<br />
• Rheowis-fluid rheometer (max. shear rate<br />
10000 s -1 )<br />
• Anton-Paar fluid rheometer<br />
• Stopped flow cell<br />
• Sample holders: 9 horizontal x 3 vertical (temperature<br />
controlled) for standard Hellma cells<br />
404.000-QX, and 110-QX<br />
• Oil & water thermostats (typical 10..100°C)<br />
• Electric thermostat (RT..200°C)<br />
• 6-positions thermostated (Peltier) sample<br />
holder (-40°C..150°C)<br />
• Magnet (horizontal, vertical)<br />
• Cryostat with sapphire windows<br />
• High temperature furnace<br />
• Pressure cells (500 bar, 2000 bar, 5000 bar)<br />
Technical Data<br />
Overall performance<br />
Q = 0.0007..0.5 Å-1 Maximal flux: 3 x 108 n/cm2 /s<br />
Typical flux: 8 x 106 n/cm2 /s<br />
(collimation 8 m, aperture 30 x 30 mm², λ = 7 Å)<br />
Velocity selector<br />
Dornier, FWHM 20 %, λ = 4.5 Å – 12 Å, 20 Å<br />
Chopper<br />
for TOF-wavelength analysis, FWHM 1 %<br />
Active apertures<br />
2 m, 4 m, 8 m, 14 m, 20 m<br />
Aperture sizes<br />
rectangular 1 x 1 mm2 – 50 x 50 mm2 Sample aperture<br />
rectangular 1 x 1 mm2 – 50 x 50 mm2 Dr. Aurel Radulescu<br />
Phone: +49.(0)89.289.10712<br />
Email: a.radulescu@fz-juelich.de<br />
Phone Instrument: .14326<br />
Figure 3: Time-resolved SANS (1s time resolution) of fast chain<br />
exchange under equilibrium between PEP(1k)-PEO(1k) cylindrical<br />
micelles using a stopped-flow apparatus at KWS-2<br />
<strong>Neutron</strong> lenses<br />
MgF , diameter 50mm, curvature 20mm<br />
2<br />
Packs with 4, 6,16 lenses<br />
Sample stage<br />
XYZθ translational-rotational stage<br />
Accuracy better than 0.01°, 0.01mm<br />
Detector 1<br />
Detection range: continuous 1 m – 20 m<br />
6Li-Scintillator 1 mm thickness + Photomultiplier<br />
Efficiency better than 95 %, Spatial resolution<br />
5.3 x 5.3 mm2 , 128 x 128 channels<br />
Max. countrate 0.6 MHz (τ = 0.64 μs)<br />
dead<br />
Detector 2 (high res.)<br />
Spatial resolution 0.45 x 0.45 mm2 Active area: Ø = 8.7 cm<br />
6Li-Scintillator 1 mm thickness<br />
Fixed position: 17 m after sample position<br />
Dr. Vitaly Pipich<br />
Phone: +49.(0)89.289.10710<br />
Email: v.pipich@fz-juelich.de<br />
www.jcns.info/jcns_kws2<br />
SANS<br />