Sample Environment<br />

• High temperature furnace<br />

• Deformation-rig with heating<br />

• Magnet (horizontal and vertical)<br />

• Sample changer with thermostat<br />

• Cryostat<br />

Technical Data<br />

Primary beam<br />

• S-shaped neutron guide (NL 4a),<br />

50 mm × 50 mm<br />

• Mechanical velocity selectors with variable<br />

speed<br />

1) ∆λ/λ = 10 % medium resolution<br />

2) ∆λ/λ = 6 % high resolution<br />

• Planned in future: chopper system with<br />

∆λ/λ = 1 %<br />

• Wavelength range: 3.5 Å – 30 Å<br />

• Wavelength resolution: 1 % - 25 % (FWHM)<br />

Polarization<br />

• Two V-shaped polarizers<br />

<strong>Source</strong>-to-sample distance<br />

• 1 m, 2 m, 4 m, 8 m, 12 m, 16 m to 20 m in<br />

steps via insertion of neutron guide sections<br />

except 1 m<br />

Dr. habil. Ralph Gilles<br />

Phone: +49.(0)89.289.14665<br />

Email: ralph.gilles@frm2.tum.de<br />

Phone Instrument: .12818 / .14992<br />

Laser System<br />

A redundant laser alignment system to position in<br />

the neutron guide, the collimation, the background<br />

apertures (BGA) or the laser in the neutron beam<br />

is installed. In addition the laser is used to support<br />

the positioning of the sample (including the sample<br />

environment) or the beam stop in a proper way.<br />

Sample size<br />

• 0 to 50 mm diameter<br />

Q-range<br />

• theoretical 0.0001 Å -1 < Q < 2 Å -1<br />

Detectors<br />

1) 128 3 He position-sensitive proportional counter<br />

with 1000 mm × 1020 mm 2 total area and 8 mm<br />

resolution and lateral detector movement up to<br />

0.5 m, counting rate up 1 MHz<br />

2) Detector with better resolution of 3 mm and<br />

0.5 m × 0.5 m total area at long distances.<br />

Dr. André Heinemann<br />

Phone: +49.(0)89.289.14534<br />

Email: andre.heinemann@hzg.de<br />

www.frm2.tum.de/sans-1<br />

SANS<br />

38 / 39