2 MEASUREMENT OF DIFFUSION COEFFICIENT<br />
2.3 MEASUREMENT OF DIFFUSION COEFFICIENT (D)<br />
As discussed above, the measurement of the diffusion coefficient can be carried out by<br />
changing either the field-gradient strength G, the duration of gradient pulse δ, or the<br />
time between two FG pulses Δ.<br />
However, in the measurement in which a time parameter such as the duration of the field<br />
gradient pulse δ or time between FG pulses Δ changes, you have to take into account<br />
the influences time, such as relaxation. Therefore, the measurement by changing<br />
magnetic-field-gradient strength G is presently in general use. The procedure for<br />
measurement of the diffusion coefficient by changing the magnetic-field-gradient<br />
strength G is explained below.<br />
■ To obtain measurement conditions<br />
1. Stop the sample spinning, and tune the probe.<br />
2. Check the 90° pulse width.<br />
In order to improve the accuracy of the diffusion coefficient measurement, we<br />
recommend that you check the 90° pulse width of the sample you are measuring.<br />
For checking the 90° pulse, perform an array measurement using the<br />
measurement mode single_pulse.ex2 or single_pulse_dec.ex2.<br />
3.<br />
4.<br />
Click on the Expmnt button in the Spectrometer Control window.<br />
The Open Experiment window opens.<br />
Click on diffusion.<br />
5.<br />
6.<br />
Select the desired sequence from the File name list.<br />
The Experiment Tool window for setting up parameters opens.<br />
Input the following value if needed.<br />
The values of x_pulse and gradient_max of the following are recalled automatically<br />
from the default values in the probe file. Input values when values obtained in<br />
step 2 differ from the values in the probe file.<br />
2-4 NM<strong>ECA</strong>/<strong>ECX</strong>-USA-3