2 MEASUREMENT OF DIFFUSION COEFFICIENT<br />

2.3 MEASUREMENT OF DIFFUSION COEFFICIENT (D)<br />

As discussed above, the measurement of the diffusion coefficient can be carried out by<br />

changing either the field-gradient strength G, the duration of gradient pulse δ, or the<br />

time between two FG pulses Δ.<br />

However, in the measurement in which a time parameter such as the duration of the field<br />

gradient pulse δ or time between FG pulses Δ changes, you have to take into account<br />

the influences time, such as relaxation. Therefore, the measurement by changing<br />

magnetic-field-gradient strength G is presently in general use. The procedure for<br />

measurement of the diffusion coefficient by changing the magnetic-field-gradient<br />

strength G is explained below.<br />

■ To obtain measurement conditions<br />

1. Stop the sample spinning, and tune the probe.<br />

2. Check the 90° pulse width.<br />

In order to improve the accuracy of the diffusion coefficient measurement, we<br />

recommend that you check the 90° pulse width of the sample you are measuring.<br />

For checking the 90° pulse, perform an array measurement using the<br />

measurement mode single_pulse.ex2 or single_pulse_dec.ex2.<br />

3.<br />

4.<br />

Click on the Expmnt button in the Spectrometer Control window.<br />

The Open Experiment window opens.<br />

Click on diffusion.<br />

5.<br />

6.<br />

Select the desired sequence from the File name list.<br />

The Experiment Tool window for setting up parameters opens.<br />

Input the following value if needed.<br />

The values of x_pulse and gradient_max of the following are recalled automatically<br />

from the default values in the probe file. Input values when values obtained in<br />

step 2 differ from the values in the probe file.<br />

2-4 NM<strong>ECA</strong>/<strong>ECX</strong>-USA-3