Excellent reproducibility using different<br />

sample injection methods<br />

Each of the three selectable sample injection methods for<br />

<strong>Autosampler</strong> A-<strong>900</strong> <strong>and</strong> A-<strong>905</strong> provide excellent sampling<br />

reproducibility:<br />

• Flushed loop: Sample loop is completely filled with<br />

sample. Sampling reproducibility using this method is<br />

better than 0.3% RSD (relative st<strong>and</strong>ard deviation).<br />

• Partial loopfill: Sample loop is partially filled with sample<br />

allowing programmable injection volumes. Partial loopfill<br />

sample injection provides sampling reproducibility better<br />

than 0.5% RSD.<br />

• µl pick-up: Sample volume is transported into the loop with<br />

transport liquid (mobile phase) after aspiration from the vial.<br />

Using this method offers minimal sample loss <strong>and</strong> maximum<br />

accuracy with a reproducibility better than 1% RSD.<br />

Different dispenser syringe sizes <strong>and</strong><br />

injection volumes provide greater flexibility<br />

Four sizes of dispenser syringes are available for<br />

<strong>Autosampler</strong> A-<strong>900</strong> <strong>and</strong> A-<strong>905</strong>: 100, 250, 500, <strong>and</strong> 1000 µl.<br />

Injection volumes for <strong>Autosampler</strong> A-<strong>900</strong> range from a<br />

few microliters to 1 ml. <strong>Autosampler</strong> A-<strong>905</strong> for ÄKTA allows<br />

injection volumes from 0.1 µl to 1 ml, making it an excellent<br />

choice for the small sample volumes injected during<br />

analytical purification.<br />

Simple programming from UNICORN<br />

<strong>Autosampler</strong> A-<strong>900</strong> is remotely controlled by UNICORN<br />

software. The UNICORN method wizard allows fast <strong>and</strong><br />

easy programming by the user <strong>and</strong> includes extensive data<br />

h<strong>and</strong>ling features (Fig 3). For <strong>Autosampler</strong> A-<strong>900</strong> <strong>and</strong><br />

A-<strong>905</strong>, an extended evaluation package, UNICORN analysis<br />

module, provides:<br />

• Quantitation by internal or external st<strong>and</strong>ards<br />

• Determination of molecular size by gel filtration<br />

• Quantitation of sample amount <strong>and</strong> sample<br />

concentration<br />

• Recovery calculations<br />

12<br />

11<br />

10<br />

9<br />

8<br />

7<br />

6<br />

5<br />

4<br />

3<br />

2<br />

1<br />

A1<br />

Multiple choice of sample vials<br />

Almost every type of vial with an outer diameter (o.d.) of<br />

12 mm or 7 mm can be used with <strong>Autosampler</strong> A-<strong>900</strong>. By<br />

using inserts, volumes of st<strong>and</strong>ard vials can be reduced.<br />

Two types of tray segments are available for <strong>Autosampler</strong><br />

A-<strong>900</strong>; st<strong>and</strong>ard trays, <strong>and</strong> microvial trays (Fig 4).<br />

<strong>Autosampler</strong> A-<strong>900</strong> is fitted with eight of the st<strong>and</strong>ard<br />

12 mm tray segments. Each segment holds 12 vials giving a<br />

maximum capacity of 96 vials. Each microvial tray segment<br />

holds twenty 7-mm vials giving a maximum capacity of<br />

160 vials.<br />

<strong>Autosampler</strong> A-<strong>905</strong> is supplied with a 96-well low microplate<br />

<strong>and</strong> 96-well high microplate for 7 mm vials, a 384-well<br />

low microplate, 48-vial adapter for 12 mm st<strong>and</strong>ard vials,<br />

<strong>and</strong> 10 ml transport solvent/wash vials (Fig 5). Samples<br />

can be directly applied to the microplates or to microvials<br />

positioned on the plates.<br />

12<br />

11<br />

10<br />

20<br />

19<br />

18<br />

17<br />

9<br />

8<br />

7<br />

16<br />

15<br />

14<br />

13<br />

6<br />

12<br />

11<br />

10<br />

9<br />

5<br />

4<br />

3<br />

8<br />

7<br />

6<br />

5<br />

2<br />

1<br />

A1<br />

4<br />

3<br />

21<br />

D1<br />

20<br />

19<br />

18<br />

17<br />

16<br />

15<br />

14<br />

13<br />

1<br />

11<br />

Fig 4. The two types of tray segments available for the A-<strong>900</strong> are (top)<br />

st<strong>and</strong>ard vials of o.d. 12 mm <strong>and</strong> (bottom) microvials of o.d. 7 mm.<br />

Fig 3. Simple programming using the method wizard of UNICORN control<br />

software.<br />

2 28-9496-70 AA

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!