Serial communication port UniNet-1<br />
Additional specifications for <strong>Autosampler</strong> A-<strong>900</strong> with cooling †<br />
Cooling capacity<br />
> 11.5°C below ambient<br />
temperature (T) for<br />
16°C < T < 40°C<br />
Typical cooling times 34 min from 23°C to 4°C<br />
(at 45% relative humidity)<br />
60 min from 32°C to 4°C<br />
(at 35% relative humidity)<br />
<strong>Autosampler</strong> A-<strong>905</strong> ‡<br />
Sampling<br />
Sample capacity<br />
96-well microplate (low/high):<br />
Direct filling or 96 microvials<br />
384-well microplate:<br />
Direct filling or 384 microvials<br />
48-vial adapter:<br />
48 vials<br />
Vial dimensions (cap included) Maximum vial height: 47 mm<br />
Minimum vial height: 12.5 mm<br />
Loop volume 5–1000 µl<br />
Dispenser syringe<br />
100 µl (st<strong>and</strong>ard in A-<strong>905</strong> for<br />
ÄKTAmicro, 250 µl, 500 µl, 1000 µl<br />
(st<strong>and</strong>ard in A-<strong>905</strong> for ÄKTA)<br />
Plate detection<br />
Missing plate detection by vial<br />
sensor<br />
Headspace pressure<br />
Built-in compressor<br />
Injection valve switch time < 100 ms<br />
Piercing precision needle ±0.6 mm<br />
Wash solvent<br />
100 ml internal wash solvent bottle<br />
Wetted parts<br />
PEEK, PTFE, glass, Tefzel, alumina,<br />
ceramic, <strong>and</strong> inert polymer<br />
Dimensions<br />
Dimensions (W ×H × D) 280 × 440 × 400 mm<br />
Weight<br />
26 kg<br />
Power requirements<br />
Power supply voltage 100–240 V, 50–60 Hz<br />
Power requirement<br />
250 VA<br />
Communication<br />
Serial communication port UniNet-1<br />
* <strong>Autosampler</strong> A-<strong>900</strong> without cooling requires UNICORN v2.3 or higher.<br />
†<br />
<strong>Autosampler</strong> A-<strong>900</strong> with cooling requires UNICORN v4.0 or higher.<br />
‡<br />
<strong>Autosampler</strong> A-<strong>905</strong> for ÄKTA requires UNICORN v4.12 or higher.<br />
Analytical performance<br />
Reproducibility<br />
Memory effect<br />
Cooling capacity<br />
< 0.3% RSD for flushed loop<br />
injections> 5 µl injection volumes<br />
< 0.5% RSD for partial loop fill<br />
injections > 5 µl injection volumes<br />
(5 µl sample, 100 µl sample loop,<br />
< 2.5 MPa)<br />
< 1.0% RSD for µl pick-up<br />
injections > 5 µl injection volumes<br />
< 0.1% with programmable<br />
needle wash<br />
Max. 12°C below ambient<br />
temperature (T) for 16°C < T < 40°C<br />
Environment<br />
Operating conditions<br />
+4°C to +40°C, 20%–95% relative<br />
humidity<br />
Sound pressure level < 70 dB (A)<br />
Installation category Category II, according to IEC 1010<br />
Pollution degree Degree 2, according to IEC 1010<br />
4 28-9496-70 AA