Communication<br />

Serial communication port UniNet-1<br />

Additional specifications for <strong>Autosampler</strong> A-<strong>900</strong> with cooling †<br />

Cooling capacity<br />

> 11.5°C below ambient<br />

temperature (T) for<br />

16°C < T < 40°C<br />

Typical cooling times 34 min from 23°C to 4°C<br />

(at 45% relative humidity)<br />

60 min from 32°C to 4°C<br />

(at 35% relative humidity)<br />

<strong>Autosampler</strong> A-<strong>905</strong> ‡<br />

Sampling<br />

Sample capacity<br />

96-well microplate (low/high):<br />

Direct filling or 96 microvials<br />

384-well microplate:<br />

Direct filling or 384 microvials<br />

48-vial adapter:<br />

48 vials<br />

Vial dimensions (cap included) Maximum vial height: 47 mm<br />

Minimum vial height: 12.5 mm<br />

Loop volume 5–1000 µl<br />

Dispenser syringe<br />

100 µl (st<strong>and</strong>ard in A-<strong>905</strong> for<br />

ÄKTAmicro, 250 µl, 500 µl, 1000 µl<br />

(st<strong>and</strong>ard in A-<strong>905</strong> for ÄKTA)<br />

Plate detection<br />

Missing plate detection by vial<br />

sensor<br />

Headspace pressure<br />

Built-in compressor<br />

Injection valve switch time < 100 ms<br />

Piercing precision needle ±0.6 mm<br />

Wash solvent<br />

100 ml internal wash solvent bottle<br />

Wetted parts<br />

PEEK, PTFE, glass, Tefzel, alumina,<br />

ceramic, <strong>and</strong> inert polymer<br />

Dimensions<br />

Dimensions (W ×H × D) 280 × 440 × 400 mm<br />

Weight<br />

26 kg<br />

Power requirements<br />

Power supply voltage 100–240 V, 50–60 Hz<br />

Power requirement<br />

250 VA<br />

Communication<br />

Serial communication port UniNet-1<br />

* <strong>Autosampler</strong> A-<strong>900</strong> without cooling requires UNICORN v2.3 or higher.<br />

†<br />

<strong>Autosampler</strong> A-<strong>900</strong> with cooling requires UNICORN v4.0 or higher.<br />

‡<br />

<strong>Autosampler</strong> A-<strong>905</strong> for ÄKTA requires UNICORN v4.12 or higher.<br />

Analytical performance<br />

Reproducibility<br />

Memory effect<br />

Cooling capacity<br />

< 0.3% RSD for flushed loop<br />

injections> 5 µl injection volumes<br />

< 0.5% RSD for partial loop fill<br />

injections > 5 µl injection volumes<br />

(5 µl sample, 100 µl sample loop,<br />

< 2.5 MPa)<br />

< 1.0% RSD for µl pick-up<br />

injections > 5 µl injection volumes<br />

< 0.1% with programmable<br />

needle wash<br />

Max. 12°C below ambient<br />

temperature (T) for 16°C < T < 40°C<br />

Environment<br />

Operating conditions<br />

+4°C to +40°C, 20%–95% relative<br />

humidity<br />

Sound pressure level < 70 dB (A)<br />

Installation category Category II, according to IEC 1010<br />

Pollution degree Degree 2, according to IEC 1010<br />

4 28-9496-70 AA