Measurement Good Practice Guide No. 93<br />
<strong>Protocol</strong> <strong>for</strong> <strong>Establishing</strong> <strong>and</strong> <strong>Maintaining</strong> <strong>the</strong> <strong>Calibration</strong> of<br />
Medical Radionuclide Calibrators <strong>and</strong> <strong>the</strong>ir Quality Control<br />
Prepared by a joint working party composed of representatives from <strong>the</strong> following:<br />
Institute of Physics <strong>and</strong> Engineering in Medicine<br />
Ionising Radiation Metrology Consultants Ltd<br />
National Physical Laboratory<br />
R GADD (1) (Chairman), M BAKER (2) (Secretary), K S NIJRAN (3), S OWENS (4),<br />
W THOMSON (5), M J WOODS (6) <strong>and</strong> F ZANANIRI (7)<br />
(1) University Hospital of North Staf<strong>for</strong>dshire, Stoke-on-Trent, Staf<strong>for</strong>dshire, ST4 7LN<br />
(2) National Physical Laboratory, Teddington, Middlesex, TW11 0LW<br />
(3) Hammersmith Hospitals NHS Trust, London, W12 0HS<br />
(4) North Western Medical Physics, Christie Hospital, Manchester, M20 4BX<br />
(5) S<strong>and</strong>well <strong>and</strong> West Midl<strong>and</strong>s Hospital Trust, Birmingham, West Midl<strong>and</strong>s, B18 7QH<br />
(6) Ionising Radiation Metrology Consultants Ltd, Teddington, Middlesex, TW11 0LW<br />
(7) Bristol General Hospital, Bristol, Avon, BS1 6SY<br />
ABSTRACT<br />
Guidance is given on <strong>the</strong> quality controls that should be undertaken on a routine basis <strong>for</strong><br />
medical radionuclide calibrators to ensure <strong>the</strong> accuracy <strong>and</strong> traceability of measurements of<br />
<strong>the</strong> activities of radiopharmaceuticals. Sources of error <strong>and</strong> magnitudes of associated<br />
uncertainties are discussed.