Coil Signature<br />

The Coil Signature is a measurement of the magnetic field strength. This value is<br />

compared to the coil signature baseline measurement taken during the flowtube<br />

signature process to determine tube calibration deviation.<br />

Electrode Circuit Resistance<br />

The Electrode Circuit Resistance is a measurement of the electrode circuit health. This<br />

value is compared to the electrode circuit resistance baseline measurement taken<br />

during the flowtube signature process to determine electrode circuit health.<br />