Specification<br />

<strong>LSM</strong> 5 <strong>EXCITER</strong> System Components<br />

Microscopes<br />

Models<br />

Z drive<br />

XY stage (option)<br />

Objectives<br />

Accessories<br />

Upright: Axio Imager.Z1m, M1m<br />

Inverted: Axio Observer.Z1m BP (base port) or SP (side port)<br />

DC motor with optoelectronic coding,<br />

smallest increment 10 or 25 nm<br />

Motorized XY scanning stage with Mark & Find (XYZ) and Tile Scan (Mosaic Scan) functions,<br />

smallest increment 1 µm<br />

More than 40 reflected-light objectives:<br />

Epiplane, Epiplan-Neofluare, EC Epiplan-Neofluare, EC Epiplan-Apochromate,<br />

LD Epiplane, LD EC Epiplan-Neofluare<br />

High-resolution AxioCam digital microscope camera<br />

Scanning Module<br />

Scanner<br />

Scanning resolution<br />

Scanning speed<br />

Two independent galvanometric scanning mirrors for rotation, zoom, offset<br />

4 x 1 to 2048 x 2048 pixels, freely selectable<br />

13 x 2 speed stages, line frequencies from 4 to 2600 Hz,<br />

5 frames/s with 512 x 512 pixels (max. 81 frames/s with 512 x 32 pixels)<br />

Scanning zoom 0.5x to 30 x, variable in digital increments of 0.1<br />

Scanning rotation Free 360° rotation, variable in increments of 0.1°<br />

Scanning field<br />

Pinhole<br />

Detection<br />

Data depth<br />

Homogeneously illuminated field of 18 mm diagonal (max.) in the primary image plane<br />

One confocal pinhole, continuous diameter variation, preadjusted<br />

One or two confocal channels (reflection/fluorescence), one optional external transmitted-light<br />

channel with DIC capability, each with high-sensitivity PMT detector<br />

Selectable between 8 bit and 12 bit<br />

Laser Modules<br />

Lasers<br />

Fiber conductor<br />

Attenuation<br />

Diode laser (405 nm) 25 mW,<br />

Ar laser (458, 488, 514 nm) 25 mW,<br />

HeNe laser (543 nm) 1 mW, HeNe laser (633 nm) 5 mW,<br />

(end-of-lifetime specification)<br />

Polarization-preserving single-mode fibers<br />

Individual and variable intensity setting of all laser lines by means of AOTF/MOTF<br />

Electronics Module<br />

<strong>LSM</strong> 5 Control<br />

Computer<br />

Controls the microscope, the laser modules, the scanning module,<br />

and further accessories; with high-performance digital signal processor (DSP)<br />

High-end PC with ample RAM and hard disk storage capacity;<br />

ergonomic high-resolution monitor or TFT flat-panel display,<br />

many accessories;<br />

Windows XP operating system with multi-user capability<br />