Sensor<br />

The sensor of the ITR 90 contains a<br />

dual filament Pirani system as well as a<br />

Bayard-Alpert measurement system.<br />

When using the degassing extension,<br />

measurements will be possible also at<br />

flange temperatures up to 150 °C<br />

10<br />

Volt<br />

8<br />

7<br />

6<br />

5<br />

4<br />

3<br />

2<br />

U(a) / Volt<br />

1<br />

0 2 468<br />

Formel Equation<br />

p = 10<br />

(u-7,75)/ 0,75<br />

10 -10 10 -9 10 -8 10 -7 10 -6 10 -5 10 -4 10 -3 10 -2 10 -1 10 0 10 mbar 1000<br />

Druck/mbar <strong>Pressure</strong><br />

Characteristic of the ITR 90<br />

58 (2.28 )<br />

67 (2.64 )<br />

Top View<br />

153 (6 )<br />

~105<br />

48 (1.89 )<br />

42 (1.65 )<br />

DN 40 CF-R<br />

DN 25 ISO-KF<br />

ø40<br />

Dimensional drawing for the ITR 90; dimensions in brackets ( ) are in inch<br />

Dimensional drawing for the degassing extension<br />

Oerlikon Leybold Vacuum Full Line Catalog<br />