01.11.2012 Views

Balloon Expandable Biliary Stent Instructions for Use - eV3

Balloon Expandable Biliary Stent Instructions for Use - eV3

Balloon Expandable Biliary Stent Instructions for Use - eV3

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Balloon</strong> <strong>Expandable</strong> <strong>Biliary</strong> <strong>Stent</strong><br />

<strong>Instructions</strong> <strong>for</strong> <strong>Use</strong>:<br />

DESCRIPTION<br />

The Intra<strong>Stent</strong> ® LD <strong>Biliary</strong> <strong>Stent</strong> is a balloon<br />

expandable stent. It is intended as a permanently<br />

implanted device. The stent is made from a 316L<br />

stainless steel tube cut into an open lattice design. It is<br />

designed to be crimped onto a non-compliant PTA<br />

balloon catheter that is indicated <strong>for</strong> biliary stent<br />

expansion. After mounting onto a balloon catheter, it is<br />

deployed and expanded by inflating the balloon.<br />

INTENTED USE<br />

The stent is intended as a palliative treatment of<br />

malignant neoplasms in the biliary tree.<br />

CONTRAINDICATIONS<br />

None known.<br />

GENERAL WARNINGS/PRECAUTIONS<br />

• The safety and effectiveness of this device <strong>for</strong> use<br />

in the vascular system have not been established.<br />

• The stent is provided STERILE <strong>for</strong> one use only<br />

and should be used by the “<strong>Use</strong> Be<strong>for</strong>e Date”<br />

printed on the package.<br />

• Carefully inspect the sterile package and stent<br />

prior to use to verify that neither has been<br />

damaged during shipment.<br />

• The inflation diameter of the balloon used during<br />

stent delivery should approximate the diameter of<br />

the obstructed duct and intended implant site.<br />

• The balloon inflation pressure should not exceed<br />

the maximum inflation pressure recommended by<br />

the manufacturer. An inflation device with a<br />

manometer is recommended.<br />

• Do not retract the balloon (catheter) unless the<br />

balloon is fully deflated under vacuum.<br />

• If resistance is encountered at any time during the<br />

insertion procedure, do not <strong>for</strong>ce passage.<br />

• If resistance occurs during movement through the<br />

sheath, the stent/delivery assembly should be<br />

withdrawn carefully.<br />

• If resistance occurs after the stent has exited the<br />

sheath or if the stent cannot be delivered to the<br />

appropriate target lesion, attempting to retract the<br />

stent/delivery assembly into the sheath may result<br />

in stent embolization. The sheath and<br />

stent/delivery assembly should be withdrawn as<br />

described below.<br />

1. Under fluoroscopic guidance, retract the<br />

stent to the exit of the sheath.<br />

2. Inflate the delivery balloon to one<br />

atmosphere to help reduce the<br />

likelihood of stent slippage or<br />

embolization.<br />

3. While preserving the guidewire position,<br />

retract the sheath and stent/delivery<br />

assembly.<br />

4. Withdraw the sheath and stent/delivery<br />

assembly together.<br />

• Recrossing of newly implanted stents (with other<br />

devices) should be done very carefully; it may<br />

cause damage to the stent architecture.<br />

• The stent may cause artifacts with MRI scans due<br />

to distortion of the magnetic field. The artifacts<br />

caused by the 316L stainless steel stent should<br />

not be greater than those caused by metal surgical<br />

clips. An MRI scan should not be per<strong>for</strong>med until<br />

the implanted stent has had a chance to heal<br />

(estimated to be 8 weeks), in order to minimize the<br />

risk of migration of the stent under a strong<br />

magnetic field.<br />

POTENTIAL HAZARDS AND SIDE<br />

EFFECTS<br />

• Infection secondary to contamination of the stent<br />

may lead to cholangitis, hemobilia, peritonitis, or<br />

abscess.<br />

• The stent may migrate from the site of implant<br />

down the biliary tract.<br />

• Overstretching of the duct may result in rupture.<br />

• Because the stent is made of 316L stainless steel,<br />

persons with known allergic reactions to 316L<br />

stainless steel may suffer an allergic response to<br />

this implant.<br />

PROCEDURE<br />

NOTE: The administration of prophylactic<br />

antibiotics may minimize the incidence of<br />

cholangitis. Procedures are tolerated well with IV<br />

analgesia.<br />

Typically this is a two-step procedure. First, the<br />

stricture is negotiated with a guidewire and an<br />

internal/external biliary drain is positioned through the<br />

stricture. Second, the transhepatic tract is accessed<br />

through the routine transhepatic approach (a lateral<br />

incision is placed between the 10 th and 11 th intercostal<br />

space, then punctured upwards under fluoroscopy<br />

aiming at the 9 th or 10 th vertebral body).<br />

• No predilation is done with malignant strictures,<br />

but a large lumen catheter may be inserted.<br />

Subsequent dilation is generally <strong>for</strong> a few seconds<br />

only.<br />

Resistance may cause damage. • Refer to pouch or outer box label <strong>for</strong> appropriate<br />

sized sheath.<br />

Page 1 45-2344-000, C FEB/05


STENT AND PTA BALLOON CATHETER<br />

ASSEMBLY PREPARATION<br />

1. Select a stent and non-compliant PTA<br />

balloon dilation catheter <strong>for</strong> the duct into<br />

which the stent will be implanted. The<br />

nominal diameter of the balloon should<br />

approximate the diameter of the obstructed<br />

duct and intended implant site.<br />

NOTE: <strong>Use</strong> a balloon that requires a 7 Fr or a<br />

larger sheath. A smaller balloon may not have<br />

a large enough diameter to allow optimal<br />

crimping of the stent to the balloon.<br />

2. Remove the stent from the package and rinse<br />

in sterile saline.<br />

3. Remove the balloon catheter from the sterile<br />

package in preparation <strong>for</strong> stent mounting.<br />

4. Preinflate the balloon to ensure the balloon<br />

opens uni<strong>for</strong>mly.<br />

5. Refold the balloon per manufacturer’s<br />

instructions.<br />

6. Visually inspect the balloon to assure that it is<br />

properly folded to its lowest profile be<strong>for</strong>e<br />

stent application.<br />

7. If balloon has a lubricious coating, use a<br />

saline saturated sterile gauze to gently wipe<br />

the folded balloon to reduce the lubricious<br />

coating.<br />

8. Carefully insert the distal tip of the<br />

unexpanded balloon, while maintaining<br />

balloon fold into the angled end of the stent<br />

mounting tube (Figure 1.)<br />

9. While holding the cap of the mounting tube,<br />

gently pinch the mounting tube behind the<br />

stent and slide the stent onto the balloon.<br />

(Figure 2). Do not compress the stent on the<br />

mounting tube. This may potentially crimp<br />

the stent to the mounting tube (Figure 2a).<br />

10. Continue to slide the stent over the balloon<br />

while using the balloon radiopaque markers<br />

<strong>for</strong> centering the balloon (Figure 3).<br />

11. Gently using the thumbs and index fingers of<br />

both hands, crimp the stent onto the balloon<br />

from all directions to ensure full adherence of<br />

the stent to the balloon.<br />

12. Flush the balloon catheter lumen with saline,<br />

then load the stent/delivery assembly onto<br />

the guidewire.<br />

13. Apply a pressure of one atmosphere to the<br />

balloon to slightly preinflate both balloon<br />

cones.<br />

14. Verify both balloon cones are slightly inflated.<br />

15. Maintain this low pressure during insertion of<br />

the stent/delivery assembly.<br />

STENT/DELIVERY ASSEMBLY<br />

INSERTION<br />

1. The stent/delivery assembly is advanced<br />

through the appropriate sized sheath, over<br />

the guidewire into correct position. If a<br />

hemostatic valve is present on the introducer<br />

sheath, a metal introducer tube can be used<br />

to facilitate advancement of the stent.<br />

2. Position the stent across the target stricture.<br />

The distal portion of the stent should be<br />

positioned so that it completely covers the<br />

distal extent of the stricture.<br />

NOTE: Care should be taken to fully cover<br />

the area proximal and distal to an inoperable<br />

malignant tumor with the stent to protect<br />

against impingement from further tumor<br />

growth.<br />

3. Pull back on the sheath to fully expose the<br />

stent. Expand the stent by inflating the<br />

balloon catheter to the nominal inflation<br />

pressure. Appropriate expansion of the<br />

balloon/stent should be determined using<br />

fluoroscopy. Do not exceed the balloon<br />

catheter manufacturer’s recommended<br />

maximum balloon inflation pressure.<br />

4. The diameter of the stent can be sized to the<br />

lumen of the bile duct.<br />

5. Deflate the balloon.<br />

6. Rotate the deflated balloon delivery catheter<br />

to ensure the stent is free and properly<br />

deployed.<br />

7. Remove the delivery catheter.<br />

Page 2 45-2344-000, C FEB/05<br />

2<br />

1<br />

2a<br />

3


PRODUCT<br />

STENT SIZE<br />

EXPANDED STENT RECOMMENDED<br />

CODE<br />

UNMOUNTED<br />

DIMENSIONS<br />

BALLOON CATHETER<br />

DIAMETER LENGTH DIAMETER LENGTH DIAMETER/LENGTH<br />

(mm) (mm) (mm) (mm)<br />

(mm)/(cm)<br />

S17-16 3.8 16.0 9.0 16.0 9.0/2.0<br />

10.0<br />

10.0/2.0<br />

12.0<br />

12.0/2.0<br />

S17-26 3.8 26.0 9.0 26.0 9.0/3.0<br />

10.0<br />

10.0/3.0<br />

12.0<br />

12.0/3.0<br />

S17-36 3.8 36.0 9.0 36.0 9.0/4.0<br />

10.0<br />

10.0/4.0<br />

12.0<br />

12.0/4.0<br />

S18-16 4.5 16.0 12.0 16.0 12.0/2.0<br />

S18-26 4.5 26.0 12.0 26.0 12.0/3.0<br />

S18-36 4.5 36.0 12.0 36.0 12.0/4.0<br />

CAUTION: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.<br />

WARRANTY DISCLAIMER<br />

Although this product has been manufactured under carefully controlled conditions, ev3 Inc. has no control over the<br />

conditions under which this product is used. ev3 Inc. there<strong>for</strong>e disclaims all warranties, both express and implied, with<br />

respect to the product including, but not limited to, any implied warranty of merchantability or fitness <strong>for</strong> a particular<br />

purpose. ev3 Inc. shall not be liable to any person or entity <strong>for</strong> any medical expenses or any direct, incidental or<br />

consequential damages caused by any use, defect, failure or malfunction of the product, whether a claim <strong>for</strong> such<br />

damages is based upon warranty, contract, tort or otherwise. No person has any authority to bind ev3 Inc. to any<br />

representation or warranty with respect to the product.<br />

The exclusions and limitations set out above are not intended to, and should not be construed so as to contravene<br />

mandatory provisions of applicable law. If any part or term of this Disclaimer of Warranty is held to be illegal,<br />

unen<strong>for</strong>ceable or in conflict with applicable law by a court of competent jurisdiction, the validity of the remaining<br />

portions of this Disclaimer of Warranty shall not be affected, and all rights and obligations shall be construed and<br />

en<strong>for</strong>ced as if this Disclaimer of Warranty did not contain the particular part or term held to be invalid.<br />

<br />

4600 Nathan Lane North<br />

Plymouth, MN 55442-2920 USA<br />

PH: 763.398.7000<br />

800.716.6700<br />

FX: 763.398.7200<br />

©2005 ev3 Inc. All rights reserved.<br />

Printed in the U.S.A.<br />

T H E E N D O V A S C U L A R C O M P A N Y<br />

Intra<strong>Stent</strong>, Mega and Max are trademarks of ev3<br />

Inc.<br />

Protected under one or more of the following: US<br />

Patent 6,827,732; 6,558,415; 6,358,274;<br />

6,254,631; 6,132,460. Non-US patents pending.<br />

Page 3 45-2344-000, C FEB/05

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!