OUTDOOR VD SERIES<br />
These models are designed to be weatherproof for indoor/outdoor use up to 1000KV DC or AC PK.<br />
They are compensated resistance-capacitance wideband DC to 1MHz and are balanced to provide<br />
safe voltage gradients with low capacitance, better high frequency accuracy (if calibrated for proximity<br />
effects), and low corona. The use of low-loss precision ceramic capacitors and exceptionally<br />
stable metal film or wire wound resistors results in stability and accuracies of 0.01% to 3% with temperature<br />
coefficients typically less than 3 PPM/°C to 40 PPM/°C and voltage coefficients from very<br />
low to zero.<br />
Applications include use with panel meters, power quality recorders, digital voltmeters, oscilloscopes<br />
or computers to measure, view and record DC, AC, transient pulse or pulse train and modulated CW.<br />
Ross Engineering Corp. dividers can be used in conjunction with Ross Engineering Corp. AC or<br />
AC/battery operated fiber optic systems or 50 or 75 ohm matching amplifier/coaxial cable drivers to<br />
create complete DC to 1MHz (or more) wideband systems for driving long fiber optic or coaxial<br />
cables. Fiber optics are particularly desirable to eliminate ground current return, pick-up and isolation<br />
problems, although Ross Engineering Corp. coaxial cable drivers provide considerable isolation<br />
between source and load also.<br />
VD900-T<br />
42.00<br />
(1066.8)<br />
INCHES<br />
(MILLIMETERS)<br />
HV TERM 3/8-24 UNF<br />
PHMS X .75(19.0)<br />
24.00<br />
(609.6)<br />
236.11 ± 2.00/VD900<br />
(5997.2) ± (50.80)<br />
258.00 ± 2.00/VD1000<br />
(6553.2) ± (50.80)<br />
HV TERM 3/8-24 UNF<br />
AS REQUIRED<br />
SEPARABLE<br />
INTO TWO OR<br />
FOUR STACKS<br />
288.0 ± 2.00<br />
(7315.2) ± (50.80)<br />
VD900- -T, BTC<br />
VD1,000- -T, BTC<br />
42.00<br />
(1066.8)<br />
HV TERM 3/8-24 UNF<br />
AS REQUIRED<br />
110.50 ± 1.00<br />
(2806.7) ± (25.4)<br />
LOW END<br />
HOUSING<br />
16.88<br />
(428.8)<br />
42.00 ± 2.00<br />
(1066.8)±(50.80)<br />
108.00 ± 2.00<br />
(2743.2) ± (50.80)<br />
VD450-<br />
-T, BTC<br />
SEPARABLE<br />
INTO TWO<br />
STACKS<br />
PRESSURE<br />
GAUGE<br />
OUTPUT<br />
TERMINALS<br />
IN WEATHER<br />
PROOF BOX<br />
PRESSURE<br />
GAUGE<br />
LOW END<br />
HOUSING<br />
OUTPUT<br />
TERMINALS<br />
IN WEATHER<br />
PROOF BOX<br />
11.50<br />
(292.1)<br />
VD60- -T<br />
VD90- -T<br />
VD120- -T<br />
32.00 ± 1.00/VD120<br />
(812.8) ± (25.4)<br />
32.00 ± 1.00/VD150<br />
(812.8) ± (25.4)<br />
58.53 ± 1.00/VD225<br />
(1486.6) ± (25.4)<br />
TANK VALVE<br />
15.50 ± 1.00/VD60-T<br />
(393.7) ± (25.4)<br />
25.00 ± 1.00/VD90-T<br />
(635.0) ± (25.4)<br />
25.00 ± 1.00/VD120-T<br />
(635.0) ± (25.4)<br />
TANK VALVE<br />
VD120-<br />
VD150-<br />
VD225-<br />
HV TERM 3/8-24 UNF<br />
PHMS X .75(19.0)<br />
-TC<br />
-BTC<br />
-BTC<br />
* ENCLOSED OPTIONAL COAXIAL OR FIBER OPTIC<br />
DRIVER / TRANSMITTER<br />
6<br />
R<br />
OSS<br />
540 Westchester Dr., Campbell, CA 95008<br />
PHONE:(408)377-4621 FAX:(408)377-5182<br />
www.rossengineeringcorp.com<br />
e-mail: info@rossengineeringcorp.com<br />
ENGINEERING CORP.