Instruction Set<br />

20<br />

What's gone wrong if …?<br />

When trouble-shooting, consult also the operating <strong>instructions</strong> for the balance.<br />

… one of the following error<br />

messages is transmitted<br />

via the interface?<br />

ES<br />

EL<br />

ET<br />

… data output is too slow/<br />

too fast?<br />

An instruction that has been received is wrong (Syntax Error); the required form of the instruction has not been<br />

observed.<br />

An instruction that has been received is wrong in content (Logistical Error).<br />

The syntax may be correct, but for some reason it cannot be executed.<br />

Example: Tare instruction given with balance overloaded or underloaded.<br />

A character sequence has not been correctly received (Transmission Error):<br />

Probably the transmission parameters of computer and balance interface do not match.<br />

The standard setting for intervals between data strings is 1 second (for GA44 Printer).<br />

The interval can be chosen when configuring (PSE):<br />

0.0, HS, 1, 2 seconds.