DCM-1021-001/00-001 version 6.0 Final<br />
3.2 RadWorks <strong>DICOM</strong> Media Server AE<br />
RadWorks provides standard conformance to the <strong>DICOM</strong> Media Storage Service<br />
and File Format (PS 3.10) and the Media Storage Application Profiles (PS 3.11) as far as the<br />
reading of uncompressed and JPEG Lossless Non-Hierarchical First Order Prediction<br />
compressed images on CD-Recordable media is concerned.<br />
3.2.1 Implementation Model<br />
3.2.1.1 Application Data Flow Diagram<br />
RWA<br />
Display<br />
Directory<br />
RWA<br />
Read<br />
Contents<br />
AE RadWorksRadWorks<br />
AE RadWorks<br />
Database Service<br />
RWA<br />
Write<br />
Contents<br />
FSC<br />
FSR<br />
Storage<br />
Medium<br />
Remote<br />
RWA<br />
The RadWorks Media Server application flow consists of the following items:<br />
• The RadWorks front-end application serves both as a FSC and a FSR. In this role it is<br />
able to read directory information from the storage medium, to import images from the<br />
storage medium and to write images and directory information to the storage medium.<br />
• The three Real World Activities involved are available through the RadWorks AE’s user<br />
interface.<br />
RadWorks does not support writing of multi session CD-R disks.<br />
3.2.1.2 File Meta Information Options<br />
The Implementation Class UID is 1.2.528.1.1001.2.20020314.1<br />
The Implementation Version Name is “AMI_<strong>DICOM</strong>00_4”<br />