08.11.2014 Views

Wide Field Camera 3 Instrument Handbook for Cycle 19 - Space ...

Wide Field Camera 3 Instrument Handbook for Cycle 19 - Space ...

Wide Field Camera 3 Instrument Handbook for Cycle 19 - Space ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

26 Chapter 4: Designing a Phase I WFC3 Proposal<br />

the background and detector sources. These sources include zodiacal light, detector<br />

dark current, and stray light from both Earth and bright targets in the field of view.<br />

Having determined the basic exposure time necessary to achieve the required S/N,<br />

you will in most cases find it necessary to achieve that total exposure time through a<br />

sequence of shorter exposures. For instance, if the exposure time is greater than the<br />

maximum orbital target visibility, it will be necessary to obtain a sequence of<br />

exposures. UVIS exposures exceeding 3,600 s require more than one exposure as do<br />

IR exposures greater than 2,800 s (see Chapter 6 and Chapter 7 <strong>for</strong> a fuller discussion).<br />

Additional reasons to structure the total exposure time are described in the<br />

following paragraphs, as well as considerations peculiar to each of the two WFC3<br />

channels.<br />

Dithering and Mosaicking<br />

A sequence of exposures obtained in a dither pattern of HST pointings will often be<br />

used to reduce the noise from flat-field calibration error, cosmic rays, and residual<br />

images. Including sub-pixel displacements in the dither pattern will allow better<br />

sampling of the point-spread function (PSF). You may design and specify a dither<br />

pattern, or use one of the pre-defined patterns already designed to sub-sample the PSF,<br />

to cover the UVIS inter-chip gap, or to mosaic a large field. The pre-defined sequences<br />

and in<strong>for</strong>mation on designing your own patterns, are presented in Appendix C of this<br />

<strong>Handbook</strong> and in the Phase II Proposal Instructions.<br />

Bright Targets<br />

For bright targets, a sequence of shorter exposures may be needed to avoid entering<br />

the non-linear or saturation regimes of the detectors (see Chapters 5, 6, and 7).<br />

Bright objects do not cause safety concerns <strong>for</strong> either UVIS or IR observations<br />

with WFC3. Image persistence can be a concern <strong>for</strong> IR observations (as discussed in<br />

Section 7.9.4 and Appendix D) but is not generally a problem with the UVIS channel.<br />

UVIS Exposures<br />

For UVIS observations, it will almost always also be desirable to use a sequence of<br />

exposures, in order to remove cosmic-ray impacts. For observations with the UVIS<br />

channel of faint targets, the effects of charge-transfer efficiency (CTE) during readout<br />

of the detector must be considered (see Chapters 5 and 6). Charge injection <strong>for</strong> such<br />

images will mitigate the non-ideal CTE, and will be offered in <strong>Cycle</strong> <strong>19</strong>.<br />

IR Exposures<br />

For observations with the IR channel you must choose a readout method from the<br />

11 available sample sequences, each of which may comprise from 1 to 15<br />

non-destructive readouts. These include RAPID (linear), SPARS (linear), and STEP<br />

(linear-log-linear) sequences (see Chapter 7). The exposure time is dictated by the<br />

sequence chosen. The ability to remove cosmic-ray impacts will depend upon the<br />

sequence chosen.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!