Conductivity - Dr. A. Kuntze GmbH
Conductivity - Dr. A. Kuntze GmbH
Conductivity - Dr. A. Kuntze GmbH
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
DR. A. <strong>Kuntze</strong><br />
product catalogue 2013<br />
<strong>Conductivity</strong><br />
valid from 01.01.2013<br />
subject to alterations
<strong>Conductivity</strong><br />
Contents<br />
Systems ....................................................................................................................................... 4<br />
SoftCon ...................................................................................................................................... 4<br />
Instruments .................................................................................................................................. 9<br />
Neon LF ...................................................................................................................................... 9<br />
K100(W)CM/ K100(W)IL ............................................................................................................. 17<br />
Sensors ...................................................................................................................................... 24<br />
Zirkon Con - 261400800 ............................................................................................................. 24<br />
LE44 Pt .................................................................................................................................... 26<br />
IL 15 ......................................................................................................................................... 30<br />
Assemblies ................................................................................................................................ 32<br />
GD 25 V(G) (PP)....................................................................................................................... 32<br />
GD 40 IL ................................................................................................................................... 34<br />
Accessories ............................................................................................................................... 37<br />
4SCR-CM ................................................................................................................................. 37<br />
4-2SCR-IL-10 ............................................................................................................................ 38<br />
Handheld Units .......................................................................................................................... 39<br />
LF 6 .......................................................................................................................................... 39
SoftCon<br />
Systems<br />
SoftCon<br />
Fast response by direct contact of the electrode to the measuring media<br />
Lowest maintenance by patented automatic sensor cleaning ASR<br />
Installation in reactor outlet<br />
Applications<br />
<strong>Dr</strong>inking Water /<br />
Beverages<br />
SoftCon<br />
Description<br />
The measuring system SoftCon is meant for conductivity measurement in the cold softening process. The complete<br />
system contains a measuring and control instrument, immersion or flow assembly, ASR (automatic sensor cleaning),<br />
sensor and cable. In comparison to the typically used pH measurement the maintenance is highly reduced by the<br />
automatic sensor cleaning.<br />
Particular characteristics<br />
Two limit values with delay, assigned to alarm relay<br />
Two separate PI-controllers<br />
Dosage check<br />
Background illuminated two-line LC-Display for measured value and temperature<br />
Scaleable, glavanically isolated 0/4.. 20 mA output<br />
Display of relay status and error messages<br />
Invertible alarm output, permanent or pulse contact<br />
OPTION: serial interface RS 485<br />
4
SoftCon<br />
Systems<br />
Block diagram<br />
Technical data<br />
Measuring parameter<br />
<strong>Conductivity</strong> (conductive)<br />
0.000.. 1.000 mS/cm<br />
0.000.. 2.000 mS/cm up to 25°C<br />
Input characteristics<br />
Accuracy<br />
+/- 0.020 mS/cm<br />
Ambient conditions Temperature 0.. 80°C<br />
5
SoftCon<br />
Systems<br />
Output characteristics<br />
Alarm relay<br />
1 relay with a potential-free NO-contact, max 250 V, 6 A, 550VA<br />
Output signal<br />
1 x 0/4.. 20 mA (scalebale, galvanically isolated)<br />
Load<br />
max. 500 Ohm<br />
Registration range Free scaleable within the measuring<br />
range<br />
Serial interface RS 485<br />
Baud rate 9600<br />
Data format 8 bit<br />
Power supply<br />
Line voltage<br />
Power consumption<br />
24/ 117/ 230 VAC, +6/-10%, 40.. 60 Hz<br />
10 VA<br />
Ambient conditions<br />
Temperature Storage -20.. +65°C<br />
Operation<br />
0.. +50°C<br />
Protection class Wall mounted IP 65<br />
Controller<br />
Control mode<br />
Relay<br />
Start delay<br />
Digital input<br />
on/off controller (adjustable hysteresis) P/PI controller (pulse-pause,<br />
pulse-frequency or continuous output)<br />
2 relays, each with a potential-free CO-contact, max 250 V, 6 A, 550VA<br />
0.. 200 sec until controller active<br />
Controller stop via external contact<br />
6
SoftCon<br />
Systems<br />
Mechanical drawing<br />
SoftCon Dip<br />
SoftCon Flow<br />
7
SoftCon<br />
Systems<br />
Assembly version<br />
Article number type/configuration Description<br />
70004210K SoftCon DIP K 100 W CM+ ASR, immersion assembly, Au-505-3O-PG<br />
70004220K SoftCon FLOW Complete measuring board with integrated ASR<br />
8
Neon LF<br />
Instruments<br />
Neon LF<br />
Easy handling by graphical menue<br />
Fitting for your application by modular setup<br />
Comfortable by firmware update and configuration via SD card<br />
Applications<br />
Process Water<br />
<strong>Dr</strong>inking Water /<br />
Beverages<br />
Waste Water Treatment<br />
Neon Touch wall mounted housing<br />
Description<br />
Neon is a modern measuring and control instrument, whose scope of functions you can determine according to your<br />
application. The basic version contains inputs for measurement and temperature, key operation, one digital input<br />
and one alarm relay. Via codes and add-on modules you can add one or two analog outputs, control functions either<br />
concentration-based or volume-based, as well as various digital output functions such as modbus interface,<br />
Ethernet, webserver, or datalogger. You can even adjust what information you want to see in the display. And if you<br />
have to install more than one instrument of the same type with the same settings, settings can easily be multiplied<br />
via SD card. Use the full advantages of the graphical menus with our touch screen version Neon Touch! Neon LF is<br />
used for conductivity measurement.<br />
9
Neon LF<br />
Instruments<br />
Particular characteristics<br />
85 .. 265 V AC power supply<br />
2 user levels by password function<br />
Safety by autolock function<br />
Power saving adjustable<br />
Alerts selectable for alarm relay<br />
SD card function: firmware update, configuration and diagnosis file<br />
Test menu for wiring check<br />
Real time clock<br />
Eventlog and event help<br />
Functional upgrade by activation code or with modules<br />
Option: 2 mA outputs 0/4 to 20 mA adjustable with fault current<br />
Option: data log function<br />
Option: RS 485 Modbus RTU<br />
Option: Ethernet with internal webserver<br />
Option: 24 V DC<br />
Option: 2nd digital input<br />
Option: 2 PID controller with digital output<br />
Option: volume-based dosage with flow measurement<br />
10
Neon LF<br />
Instruments<br />
Block diagram<br />
Technical data<br />
Measuring parameter<br />
<strong>Conductivity</strong> (inductive)<br />
<strong>Conductivity</strong> (conductive)<br />
0.. 2.000 mS/cm<br />
0.. 20.00 mS/cm<br />
0.. 200.0 mS/cm<br />
0.. 2000 mS/cm<br />
0.. 2.000 µS/cm C=0,05 /cm<br />
0.. 20.00 µS/cm C=0,05 /cm<br />
0.. 200.0 µS/cm C=0,05 /cm<br />
0.. 2.000 mS/cm C=0,20 /cm<br />
0.. 20.00 mS/cm C=1,00 /cm<br />
0.. 200.0 mS/cm C=10 /cm<br />
11
Neon LF<br />
Instruments<br />
Input characteristics<br />
Temperature measuring range -30.0 .. +140.0 °C<br />
Temperature coefficient<br />
0.0 .. 8.0 %/K adjustable or non-linear<br />
Digital input<br />
1 as controller stop by external contact; option: 2nd as controller stop or flow<br />
measurement for volume based dosing<br />
Output characteristics<br />
Alarm relay<br />
Output signal<br />
Storage media<br />
Serial interface<br />
1 potential-free NO contact, max. 250 V, 6 A, 550 VA (invertable)<br />
Option: 2 x 0/4 .. 20 mA (scaleable, galvanically isolated)<br />
Load<br />
500 Ohm<br />
Registration range Scaleable within the measuring range<br />
Accessorie: SD card up to 2 MB - industrial standard<br />
Option: RS 485 Modbus RTU<br />
Baud rate<br />
19200 bps<br />
Data format 8 bit<br />
Ethernet<br />
Option: internal webserver<br />
Power supply<br />
Line voltage<br />
Power consumption<br />
85.. 250 V AC, +6/-10%, 40.. 60Hz option: 24 V DC<br />
10 VA<br />
Ambient conditions<br />
Temperature Storage -20.. +65°C<br />
Operation<br />
0 .. +50°C<br />
Humidity<br />
max. 90% rH at 40°C (non-condensing)<br />
Protection class Wall mounted IP 65<br />
Panel mounted IP 54 (front), IP 30 (housing)<br />
Controller<br />
Control response<br />
Relay<br />
Start delay<br />
Controller stop<br />
Option: on/off controller (adjustable hysteresis) P/PI/ PID controller<br />
(pulse-pause, pulse-frequency or continuous output) servo motor control<br />
2 relays, each with a potential-free NO contact, max. 250V, 6 A, 550 VA<br />
0.. 200 sec until controller active<br />
Digital input<br />
Proportion to volume<br />
Control mode<br />
Option: quantity based by flow measurement<br />
Flow measurement impuls measurement NPN (by digital input 2)<br />
Flow measurement Engine speed 0,030.. 9,999 l/Imp<br />
Relay 1<br />
Potential-free NO contact, max. 250V, 6 A, 550 VA output of<br />
control variable as pulse frequency or pulse pause.<br />
Relay 2<br />
Activating circulation pump<br />
12
Neon LF<br />
Instruments<br />
Certificates and approvals<br />
CE-Symbol<br />
The product meets the requirements of the harmonized European standards.<br />
It thus complies with the legal requirements of the EC directives.<br />
EMV/EMC EN 61000 6-1 (3) EN 61000 6-2 (4) EN 61326-1<br />
Design configuration<br />
Material<br />
Dimensions<br />
Mounting dimension<br />
Weight<br />
Connection<br />
ABS<br />
Panel mounted housing: 138x138x83 mm; Wall mounted housing:<br />
144x144x156 mm<br />
Panel mounted housing: 138x138x42 mm<br />
0.6 kg (wall mounted housing: 1 kg)<br />
Cable inlet: 2x M16, 2x M12 + optional: 2x M12 and 1x M25 plug-in terminal:<br />
rigid /flexible 0.2-2.5 mm² / 0.2-2.5 mm² measurement rigid /flexible 0.2-1<br />
mm² / 0.2-1.5 mm²<br />
Mechanical drawing<br />
Panel mounted housing<br />
13
Neon LF<br />
Instruments<br />
Wall mounted housing<br />
14
Neon LF<br />
Instruments<br />
Order information<br />
Choose the components you need and that´s how your "assembly version" is designed. We will have to technically<br />
inspect and approve a free combination of individual key features.<br />
15
Neon LF<br />
Instruments<br />
Accessories<br />
accessories<br />
Measuring point setup<br />
industrial-SD-card<br />
16
K100(W)CM/ K100(W)IL<br />
Instruments<br />
K100(W)CM/ K100(W)IL<br />
Easy and save operation by plain text menu guidance<br />
Safety by password function<br />
Linear and non-linear special pure water temperature compensation<br />
Applications<br />
Cooling water<br />
Waste Water Treatment<br />
Process Water<br />
K 100 W CM<br />
Description<br />
The K 100 (W) CM and the K 100 (W) IL are sophisticated single-channel instruments for measuring and control of<br />
conductive and inductive conductivity. Temperature measurement of the measuring media is integrated. The<br />
instruments are available in a panel mounted housing (K100 CM resp. IL) and in a wall mounted housing (K 100 W<br />
CM resp. IL).<br />
Particular characteristics<br />
2 limit values with delay, assigned to alarm relay<br />
2 separate PI-controllers<br />
Dosage check<br />
Background illuminated two-line LC-Display for measured value and temperature<br />
Scaleable, glavanically isolated 0/4.. 20 mA output<br />
Display of relay status and error messages<br />
Invertible alarm output, permanent or pulse contact<br />
Option: serial interface RS 485<br />
17
K100(W)CM/ K100(W)IL<br />
Instruments<br />
Block diagram<br />
Technical data<br />
Measuring parameter<br />
<strong>Conductivity</strong> (conductive)<br />
<strong>Conductivity</strong> (inductive)<br />
0.. 2.000 µS/cm c= 0.05 /cm<br />
0.. 20.00 µS/cm c= 0.05 /cm<br />
0.. 200.0 µS/cm c= 0.05 /cm<br />
0.. 2.000 mS/cm c= 0.20 /cm<br />
0.. 20.00 mS/cm c= 1.00 /cm<br />
0.. 200.0 mS/cm c= 10 /cm<br />
0.. 2.000 mS/cm<br />
0.. 20.00 mS/cm<br />
0.. 200.0 mS/cm<br />
0.. 2000 mS/cm<br />
18
K100(W)CM/ K100(W)IL<br />
Instruments<br />
Input characteristics<br />
Temperature measuring range<br />
Temperature coefficient<br />
Digital input<br />
-30.. +140°C (display range)<br />
0.. 8.0 %/°K adjustable or non-linear<br />
1 as controller stop by external contact<br />
Output characteristics<br />
Alarm relay<br />
1 potential-free NO contact, max. 250 V, 6 A, 550 VA<br />
Output signal<br />
1 x 0/4.. 20 mA (scaleable, galvanic isolated)<br />
Load<br />
max. 500 Ohm<br />
Registration range Scaleable within the measuring range<br />
Serial interface Option: RS 485<br />
Baud rate<br />
9600 baud<br />
Data format 8 bit<br />
Power supply<br />
Line voltage<br />
Power consumption<br />
24 / 117/ 230 VAC, +6/-10%, 40.. 60 Hz<br />
10 VA<br />
Ambient conditions<br />
Temperature Storage -20.. +65°C<br />
Operation<br />
0.. +50°C<br />
Humidity<br />
max. 90% rH at 40°C (non-condensing)<br />
Protection class Wall mounted IP 65<br />
Panel mounted IP 54 (front) IP 30 (housing)<br />
Controller<br />
Control response<br />
Relay<br />
Start delay<br />
Controller stop<br />
On/off controller (adjustable hysteresis) P/PI controller (pulse-pause,<br />
pulse-frequency or continuous output)<br />
2 relays each with potential-free NO contact, max. 250 V, 6 A, 550 VA<br />
0.. 200 sec until controller active<br />
digital input<br />
Certificates and approvals<br />
CE-Symbol<br />
EMV/EMC<br />
The product meets the requirements of the harmonized European standards.<br />
It thus complies with the legal requirements of the EC directives.<br />
EN 61326 class B, 2004/108/EG<br />
Design configuration<br />
Material<br />
Dimensions<br />
Mounting dimension<br />
Weight<br />
Panel mounted housing: Noryl; Wall mounted hosuing: ABS<br />
Panel mounted housing: 90x90x116 mm; Wall mounted housing: 165x160x85<br />
mm<br />
Panel mounted housing: 92x92x140 mm<br />
Panel mounted housing: 0,75 kg; Wall mounted housing: 0,95 kg<br />
19
K100(W)CM/ K100(W)IL<br />
Instruments<br />
Mechanical drawing<br />
Wall mounted housing<br />
Panel mounted housing<br />
20
K100(W)CM/ K100(W)IL<br />
Instruments<br />
Storage version<br />
Article number type/configuration Description<br />
100100K K 100 CM Conductive conductivity, panel mounted, 230 V AC<br />
100190K K 100 IL Inductive conductivity, panel mounted, 230 V AC<br />
105100K K 100 W CM Conductive conductivity, wall mounted, 230 V AC<br />
105190K K 100 W IL Inductive conductivity, wall mounted, 230 V AC<br />
21
K100(W)CM/ K100(W)IL<br />
Instruments<br />
Accessories<br />
hand-held units<br />
LF 6<br />
All-purpose portable instrument for measurement of<br />
conductivity and temperature. Sensor with integrated<br />
temperature sensor is attached. Sensor, instrument<br />
and portable case are included in delivery.<br />
accessories<br />
S-341 data logger The bus data logger S-341 on the base of the RS-485<br />
with <strong>Kuntze</strong> protocol reads, shows, and stores<br />
measuring data of up to 10 <strong>Kuntze</strong> instruments with<br />
up to 8 parameters each.<br />
Measuring point setup<br />
inductive measurement<br />
22
K100(W)CM/ K100(W)IL<br />
Instruments<br />
conductive measurement<br />
23
Zirkon Con - 261400800<br />
Sensors<br />
Zirkon Con - 261400800<br />
Easy installation by 12 mm sensor design<br />
Integrated temperature sensor<br />
5 m fixed cable<br />
Applications<br />
<strong>Dr</strong>inking Water /<br />
Beverages<br />
Process Water<br />
Description<br />
Zirkon Con 261400800<br />
2-electrode conductivity sensor for electrolytic conductivity with integrated temperature sensor Pt100. The body<br />
material is Epoxy, the electrodes are made of graphite.<br />
24
Zirkon Con - 261400800<br />
Sensors<br />
Technical data<br />
Measuring parameter<br />
<strong>Conductivity</strong> (conductive)<br />
0.00.. 20.00 mS/cm<br />
Ambient conditions<br />
Max. pressure < 2 bar at 25°C<br />
Temperature 0.. 80.0 °C<br />
Mechanical construction<br />
Shaft material<br />
Epoxy<br />
Mechanical Construction<br />
120 mm<br />
Electrode material<br />
graphite<br />
Mechanical connection revolving PG 13,5<br />
Electrical connection<br />
5 m fixed cable<br />
Temperature sensor Pt 100<br />
Mechanical drawing<br />
261400800<br />
Storage version<br />
Article number type/configuration Description<br />
26120004K 261400800 Conducitivity sensor, c = 1,0 mS/cm<br />
25
LE44 Pt<br />
Sensors<br />
LE44 Pt<br />
Temperature sensor integrated<br />
Two versions: 4-pole angular plug (Hirschmann) with 3/4'' male thread or 10 m cable with<br />
1/2'' male thread<br />
Little dirt-prone and easy to clean<br />
Applications<br />
Cooling water<br />
<strong>Dr</strong>inking Water /<br />
Beverages<br />
Process Water<br />
LE44 Pt<br />
LE44 Pt 4SCR<br />
Description<br />
2-electrode conductivity sensor for electrolytic conductivity with integrated temperature sensor Pt100. The body<br />
material is PVDF, the electrodes are made of stainless steel (concentric electrodes, material number 1.4571). The<br />
sensor is a available as flow version with male thread 3/4" and 4-pole angular plug connector (Hirschmann) for<br />
electrical connection or as immersion version with 10 m fixed cable and 1/2'' male thread.<br />
26
LE44 Pt<br />
Sensors<br />
Technical data<br />
Measuring parameter<br />
<strong>Conductivity</strong> (conductive)<br />
0.000.. 2.000 µS/cm c= 0.05 /cm<br />
0.00.. 20.00 µS/cm c= 0.05 /cm<br />
0.0.. 200.0 µS/cm c= 0.05 /cm<br />
0.000.. 2.000 mS/cm c= 0.20 /cm<br />
0.00.. 20.00 mS/cm c= 1.00 /cm<br />
Ambient conditions<br />
Max. pressure<br />
16 bar (at 25°C)<br />
Plug connection Cable connection<br />
Temperature 0.. 135 °C 0.. 90 °C<br />
Mechanical construction<br />
Shaft material<br />
PVDF - plug version, PP - cable version<br />
Electrode material Stainless steel 1.4571<br />
Mechanical connection<br />
3/4'' male thread - plug version, 1/2'' male thread - cable version<br />
Electrical connection<br />
4-pole angular plug (Hirschmann), 10 m connected cable<br />
Temperature sensor Pt 100<br />
27
LE44 Pt<br />
Sensors<br />
Mechanical drawing<br />
LE44 Pt<br />
LE44 Pt 4SCR<br />
28
LE44 Pt<br />
Sensors<br />
Storage version<br />
Article number type/configuration Description<br />
26125252K LE44 Pt, c=0,05 <strong>Conductivity</strong> sensor (0.. 2; 0.. 20; 0.. 200 µS/cm)<br />
26125253K LE44 Pt, c=0,20 <strong>Conductivity</strong> sensor (0.. 2 mS/cm)<br />
26125254K LE44 Pt, c=1,00 <strong>Conductivity</strong> sensor (0.. 20 mS/cm)<br />
26125257K LE44 Pt 4SCR, <strong>Conductivity</strong> sensor (0.. 2; 0.. 20; 0.. 200 µS/cm) with 10 m fixed cable<br />
c=0,05<br />
26125255K LE44 Pt 4SCR, <strong>Conductivity</strong> sensor (0.. 2 mS/cm) with 10 m fixed cable<br />
c=0,20<br />
26125255K LE44 Pt 4SCR,<br />
c=1,00<br />
<strong>Conductivity</strong> sensor (0.. 20 mS/cm) with 10 m fixed cable<br />
29
IL 15<br />
Sensors<br />
IL 15<br />
Corrosion resistant<br />
Dirt resistant<br />
Small robust construction<br />
Applications<br />
Process Water<br />
Cooling water<br />
Waste Water Treatment<br />
IL 15<br />
Description<br />
<strong>Conductivity</strong> sensor for the inductive conductivity measurement in Polypropylene (PP) with two internal, circular<br />
measuring coils. The sensor is delivered with 6 m cable and internal temperature sensor.<br />
30
IL 15<br />
Sensors<br />
Technical data<br />
Measuring parameter<br />
<strong>Conductivity</strong> (inductive)<br />
0.000.. 2.000 mS/cm<br />
0.00.. 20.00 mS/cm<br />
0.0.. 200.0 mS/cm<br />
0.. 2000 mS/cm<br />
Ambient conditions<br />
Max. pressure<br />
10 bar (at 20 °C)<br />
Mechanical construction<br />
Shaft material<br />
Mechanical connection<br />
Temperature sensor<br />
PP<br />
1/2'' male thread<br />
NTC<br />
Mechanical drawing<br />
IL 15<br />
Storage version<br />
Article number type/configuration Description<br />
26164025K IL 15 Inductive conductivity sensor, incl. 6 m cabel<br />
31
GD 25 V(G) (PP)<br />
Assemblies<br />
GD 25 V(G) (PP)<br />
Easy to dismantle<br />
Easy mounting by bonding or 1'' female thread<br />
GD 25 VG<br />
Description<br />
Flow assembly for installation of one conductive sensor LE 44 Pt in pipes with adhesive coupling or pipe coupling<br />
DN 25 with 1" female thread. Available in PVC and PP.<br />
Technical data<br />
Ambient conditions<br />
Max. pressure<br />
Temperature<br />
PVC: 16 bar (at 20°C) PP: 10 bar (at 20°C)<br />
max. 40°C (PVC) max. 90°C (PP)<br />
Mechanical construction<br />
Material<br />
PVC, PP<br />
Dimensions<br />
see dimension drawing<br />
Installation GD 25 V: adhesive coupling GD 25 VG (PP): pipe coupling DN 25 with 1"<br />
female threat<br />
32
GD 25 V(G) (PP)<br />
Assemblies<br />
Mechanical drawing<br />
GD 25 V(G)<br />
product variants<br />
Article number type/configuration Description<br />
36604230K GD 25 V adhesive coupling (DN 25), PVC<br />
36604231K GD 25 VG pipe coupling (DN 25) with 1'' internal thread, PVC<br />
36604235K GD 25 VG PP pipe coupling (DN 25) with 1'' internal thread, PP<br />
33
GD 40 IL<br />
Assemblies<br />
GD 40 IL<br />
Installation of the sensor by threaded connection<br />
For measuring in pipes<br />
GD 40<br />
Description<br />
PP or PVC Flow assembly for installation of one conductivity sensor IL 15.<br />
Particular characteristics<br />
Installation of the sensor by threated connection<br />
For measuring in pipes<br />
Information<br />
Picture shows the PP version of the product.<br />
Technical data<br />
Ambient conditions<br />
Max. pressure<br />
PVC: 16 bar (at 20°C) PP: 10 bar (at 20°C)<br />
Temperature PVC: max.40°C PP: max. 90°C<br />
34
GD 40 IL<br />
Assemblies<br />
Mechanical construction<br />
Material<br />
Dimensions<br />
Installation<br />
PP, PVC<br />
see dimension drawing<br />
PVC: adhesive coupling d=50 mm PP: welded sleeve d=50 mm<br />
Mechanical drawing<br />
GD 40 IL<br />
35
GD 40 IL<br />
Assemblies<br />
product variants<br />
Article number type/configuration Description<br />
36604026K GD 40 IL PVC adhesive coupling d= 50 mm, PVC<br />
36604025K GD 40 IL PP welded sleeve d=50mm, PP<br />
36
4SCR-CM<br />
Accessories<br />
4SCR-CM<br />
Low cable capacitance<br />
Screened<br />
Sold by meter<br />
Description<br />
Connecting cable for conductive conductivity sensors type LE 44 Pt.<br />
Technical data<br />
Ambient conditions<br />
Ambient temperature<br />
-20 .. +70°C<br />
Mechanical construction - panel mounted housing<br />
Material<br />
PVC<br />
product variants<br />
Article number type/configuration Description<br />
44136352K 4-SCR-CM connection cable conductivity sensors, sold by meter<br />
37
4-2SCR-IL-10<br />
Accessories<br />
4-2SCR-IL-10<br />
Screened<br />
Description<br />
Extension cable - 10 meter - for the inductive conductivity sensor IL 15<br />
Technical data<br />
Ambient conditions<br />
Ambient temperature<br />
-20 .. +70°C<br />
Mechanical construction - panel mounted housing<br />
Material<br />
PVC<br />
product variants<br />
Article number type/configuration Description<br />
44136353K 4-2SCR-IL-10 10 m Extension cable for sensor IL15<br />
38
LF 6<br />
Handheld Units<br />
LF 6<br />
Automatic or manual temperature compensation<br />
4 measuring ranges with automatic changeover<br />
min/max value memory, hold function<br />
LF 6 Set<br />
Description<br />
All-purpose portable instrument for measurement of conductivity and temperature. Sensor with integrated<br />
temperature sensor is attached. Sensor, instrument and portable case are included in delivery.<br />
Particular characteristics<br />
Automatic measuring-value-stability recognition<br />
Automatic power shut-off<br />
Double display for conductivity value and temperature<br />
Power supply by battery or mains operation<br />
Easy and safe handling by soft keys<br />
Low battery indication<br />
With fold away support<br />
Technical data<br />
Measuring parameter<br />
<strong>Conductivity</strong> (conductive)<br />
0.. 200.0 µS/cm<br />
0.. 2000 µS/cm<br />
0.. 20.00 mS/cm<br />
0.. 200.0 mS/cm<br />
39
LF 6<br />
Handheld Units<br />
Power supply<br />
Line voltage<br />
Power supply<br />
Power consumption<br />
via 10-12 V (power unit not included in delivery)<br />
9V battery<br />
ca. 3 mA<br />
Ambient conditions<br />
Ambient temperature operation 0 .. 50°C storage -20° .. 70°C<br />
Relative humidity<br />
max. 90% rH at 40°C (non-condensing)<br />
Protection class IP 65<br />
Certificates and approvals<br />
CE-Symbol<br />
Declaration of conformity: The product meets the requirements of the<br />
harmonized European standards. It thus complies with the legal requirements<br />
of the EC directives. The manufacturer confirms successful testing of the<br />
product by affixing the CE symbol.<br />
product variants<br />
Article number type/configuration Description<br />
65522410K LF 6 complete set measuring instrument with sensor including portable case<br />
40
The <strong>Dr</strong>. A. <strong>Kuntze</strong> <strong>GmbH</strong> "Terms and Conditions" will be applicable.<br />
<strong>Conductivity</strong><br />
Notes
<strong>Conductivity</strong><br />
Notes<br />
The <strong>Dr</strong>. A. <strong>Kuntze</strong> <strong>GmbH</strong> "Terms and Conditions" will be applicable.
The <strong>Dr</strong>. A. <strong>Kuntze</strong> <strong>GmbH</strong> "Terms and Conditions" will be applicable.<br />
<strong>Conductivity</strong><br />
Notes
Powered by TCPDF (www.tcpdf.org)<br />
<strong>Dr</strong>. A. <strong>Kuntze</strong> <strong>GmbH</strong><br />
Robert-Bosch-Str. 7a | D-40668 Meerbusch | Germany<br />
Fon +49 (0) 2150 70 66 -0 | Fax +49 (0) 2150 70 66 -60<br />
info@kuntze.com | www.kuntze.com