7-8<br />
7.5 Alarm Summary Tables<br />
The following tables summarize the patient, system, and power alarm information.<br />
Patient Alarms<br />
Alarm<br />
LED Indicator<br />
Display<br />
Message<br />
Audible<br />
Indicator<br />
Device Action Possible Cause Patient Action Provider Action<br />
Apnea<br />
Red flash<br />
ALARM and<br />
APNEA<br />
flash<br />
• • • • •<br />
Operates<br />
Apnea event<br />
occurred<br />
during therapy.<br />
Press the RESET<br />
button to reset<br />
the alarm.<br />
Report the alarm<br />
to your health care<br />
professional.<br />
Continue using<br />
your device.<br />
Verify patient<br />
status.<br />
Patient<br />
Disconnect<br />
Red flash<br />
ALARM and<br />
PATIENT<br />
flash<br />
• • • • •<br />
Operates<br />
Patient circuit is Press the RESET<br />
disconnected or button to reset<br />
has a large leak. the alarm.<br />
Reconnect the<br />
patient circuit or<br />
correct the leak.<br />
If the alarm<br />
continues, contact<br />
your health care<br />
professional.<br />
Reconnect the<br />
patient circuit or<br />
fix the leak.<br />
If the alarm<br />
continues, contact<br />
an authorized<br />
service<br />
representative or<br />
<strong>Respironics</strong> to<br />
have the device<br />
serviced.<br />
Low<br />
Minute<br />
Ventilation<br />
Red flash<br />
ALARM and<br />
MinVent<br />
flash<br />
• • • • •<br />
Operates<br />
The calculated<br />
minute<br />
ventilation is<br />
less than or<br />
equal to the<br />
alarm setting.<br />
Press the RESET<br />
button to reset<br />
the alarm.<br />
Report the alarm<br />
to your health care<br />
professional.<br />
Continue using<br />
your device.<br />
Verify patient<br />
status.<br />
Provider Manual