12-3<br />
Triggers and Cycles<br />
Patient Inspired Volume:<br />
Spontaneous Trigger:<br />
6 ml<br />
• Shape Trigger<br />
• Volume 6 ml above Vleak<br />
Spontaneous Cycle:<br />
• Spontaneous Expiratory Threshold (SET)<br />
• IPAP Maximum of 3.0 seconds<br />
Connector<br />
The patient interface port is a 22 mm tapered connector.<br />
Disposal<br />
Dispose of this device in accordance with local regulations.<br />
Pressure Drop Versus Flow for Patient Circuits<br />
The device automatically compensates for pressure drops associated with a 6-foot (182.9 cm) smooth bore tube. Additional<br />
pressure drops will occur when restrictive elements are added to the patient circuit. The following graph shows the additional<br />
pressure drop when adding:<br />
1. A bacteria filter<br />
2. A bacteria filter and a <strong>Respironics</strong> Pass-over humidifier<br />
NOTE:<br />
Always use a manometer to verify patient mask pressure.<br />
Pressure Drop Versus Flow for Patient Circuits<br />
Bacteria Filter Bacteria Filter and <strong>Respironics</strong> Passover Humidifier<br />
10<br />
9<br />
8<br />
7<br />
6<br />
5<br />
4<br />
3<br />
2<br />
1<br />
0<br />
0 20 40 60 80 100 120 140 160 180<br />
Flow LPM<br />
Figure 12–1 Pressure Drop Versus Flow for Patient Circuits<br />
Provider Manual