BIOLOGIC CALIBRATION CHECK<br />
(Optional if Threshold Audiometry is not performed)<br />
To insure that the audiometer is in calibration, the person with primary responsibility for the<br />
hearing screening program should do biologic calibration checks. These calibration checks<br />
should be done each day prior to use or anytime during use when there is reason to suspect the<br />
audiometer may not be working properly. Use the “Same Ear HL” procedure described below.<br />
(Form on next page.)<br />
Same Ear HL Procedure:<br />
1. Obtain a threshold (500–8000 Hz) on the better ear with right (red) earphone and record<br />
the results.<br />
2. Obtain a threshold (500–8000 Hz) on the same ear with the left (blue) earphone and<br />
record the results.<br />
3. Check the thresholds at each frequency to see that they differ by no more that ±5 dB. If<br />
they do vary more than this, the audiometer should not be used and the audiometer should<br />
be checked.<br />
35