23.11.2014 Views

Multiparameter Photometer - Comlibris

Multiparameter Photometer - Comlibris

Multiparameter Photometer - Comlibris

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

SPECIFICATIONS<br />

HI 96 series<br />

Water Resistant Single Parameter <strong>Photometer</strong>s<br />

with HANNA’s Exclusive CAL CHECK<br />

Parameter Code* Range Precision<br />

Ammonia,<br />

Medium Range<br />

Chlorine,<br />

Ultra High Range<br />

HI 96715<br />

HI 96771<br />

0.00 to 9.99 mg/L<br />

(as NH 3 -N)<br />

UHR: 0 to 500 mg/L<br />

LR: 0.00 to 5.00 mg/L<br />

Chlorine, Free HI 96701 0.00 to 5.00 mg/L<br />

Chlorine, Free<br />

for Drinking Water Applications<br />

HI 96762<br />

Chlorine, Free and Total HI 96781<br />

Chlorine, Free and Total HI 96724<br />

Chlorine, Free and Total,<br />

High Range<br />

Chlorine, Total for the Analysis of<br />

Trace Total Chlorine Concentration<br />

Copper,<br />

Low Range<br />

HI 96734<br />

HI 96761<br />

HI 96747<br />

Detergents, Anionic HI 96769<br />

Fluoride,<br />

Low Range<br />

Iron,<br />

High Range<br />

Nitrite,<br />

Low Range<br />

Phosphate,<br />

High Range<br />

HI 96729<br />

HI 96721<br />

HI 96707<br />

HI 96717<br />

0.000 to 0.500 mg/L<br />

Free Cl 2 : 0.00 to 5.00 mg/L;<br />

Total Cl 2 : 0.00 to 5.00 mg/L<br />

Free Cl 2 : 0.00 to 5.00 mg/L;<br />

Total Cl 2 : 0.00 to 5.00 mg/L<br />

Free Cl 2 : 0.00 to 10.00 mg/L;<br />

Total Cl 2 : 0.00 to 10.00 mg/L<br />

0.000 to 0.500 mg/L<br />

0.000 to 1.500 mg/L<br />

0.00 to 3.50 mg/L<br />

(as SDBS)<br />

0.00 to 2.00 mg/L<br />

0.00 to 5.00 mg/L<br />

0.000 to 0.600 mg/L<br />

(as NO 2<br />

- -N)<br />

0.0 to 30.0 mg/L<br />

Phosphorus HI 96706 0.0 to 15.0 mg/L<br />

Silica HI 96770 0 to 200 mg/L (as SiO 2 )<br />

±0.12 mg/L<br />

@ 6.00 mg/L<br />

±2 mg/L<br />

@ 100 mg/L<br />

±0.02 mg/L<br />

@ 1.00 mg/L<br />

±0.02 mg/L<br />

@ 1.00 mg/L<br />

±0.004 mg/L<br />

@ 0.200 mg/L<br />

±0.02 mg/L<br />

@ 1.00 mg/L<br />

±0.03 mg/L<br />

@ 1.00 mg/L<br />

±0.06 mg/L<br />

@ 3.00 mg/L<br />

±0.004 mg/L<br />

@ 0.200 mg/L<br />

±0.015 mg/L<br />

@ 0.750 mg/L<br />

±0.04 mg/L<br />

@ 1.00 mg/L<br />

±0.03 mg/L<br />

@ 1.00 mg/L<br />

±0.01 mg/L<br />

@ 1.50 mg/L<br />

±0.001 mg/L<br />

@ 0.100 mg/L<br />

±0.5 mg/L<br />

@ 12 mg/L<br />

±0.2 mg/L<br />

@ 6.0 mg/L<br />

±5 mg/L<br />

@ 100 mg/L<br />

Narrow<br />

Band Filter<br />

Method<br />

@ 466 nm<br />

Adaptation of the ASTM Manual of Water and<br />

Environmental Technology, D1426-93, Nessler Method.<br />

@ 525 nm<br />

@ 525 nm<br />

@ 525 nm<br />

@ 525 nm<br />

@ 525 nm<br />

@ 525 nm<br />

@ 525 nm<br />

@ 560 nm<br />

@ 610 nm<br />

@ 575 nm<br />

@ 525 nm<br />

@ 525 nm<br />

@ 525 nm<br />

@ 525 nm<br />

@ 466 nm<br />

Adaptation of Standard Methods, 20th edition, 4500-Cl.<br />

Adaptation of the USEPA method 330.5 and Standard<br />

Method 4500-Cl G.<br />

Adaptation of Standard Methods, 20th edition,<br />

4500-Cl G.<br />

Adaptation of the USEPA method 330.5 and Standard<br />

Method 4500-Cl G.<br />

Adaptation of the USEPA method 330.5 and Standard<br />

Method 4500-Cl G.<br />

Adaptation of the USEPA method 330.5 and Standard<br />

Method 4500-Cl G.<br />

Adaptation of Standard Methods, 20th edition,<br />

4500-Cl G.<br />

Adaptation of the USEPA approved bicinchoninate<br />

method.<br />

Adaptation of the USEPA method 425.1 for drinking<br />

waters, surface waters, domestic and industrial wastes<br />

and Standard Methods, 20th edition, 5540C, Anionic<br />

Surfactants as MBAS.<br />

Adaptation of the EPA method 340.1 and SPADNS<br />

method.<br />

Adaptation of the USEPA method 315B and Standard<br />

Method 3500-Fe B.<br />

Adaptation of an EPA approved method.<br />

Amino Acid Method, adapted from Standard Method<br />

for the Examination of Water and Wastewater.<br />

Amino Acid Method, adapted from Standard Method<br />

for the Examination of Water and Wastewater.<br />

Adaptation of the USEPA method 370.1 for drinking,<br />

surface and saline waters, domestic and industrial<br />

wastes and Standard Method 4500-SiO 2 C.<br />

Single Parameter <strong>Photometer</strong>s<br />

Portable Meters<br />

SPECIFICATIONS common to all models<br />

Light Source<br />

Tungsten lamp; LED (HI 96715 and HI 96770 only)<br />

Environment<br />

0 to 50°C (32 to 122°F); Max 95% RH non-condensing<br />

Battery Type/Life<br />

(1) 9V/approx. 200 hours continuous use<br />

Auto Shut-off<br />

After 10 minutes of non-use in measurement mode<br />

Dimensions/Weight<br />

192 x 102 x 67 mm (7.6 x 4 x 2.6”)/290 g (10 oz.)<br />

ORDERING INFORMATION<br />

HI 96 series of photometers are supplied with<br />

sample cuvets with caps, 9V battery and<br />

instruction manual.<br />

HI 96 “C” series of photometers is supplied with<br />

CAL CHECK standard cuvets, sample cuvets with<br />

caps, 9V battery, cuvet cleaning cloth and<br />

instruction manual in a hard carrying case<br />

(additional accessories such as glass vials, pipettes<br />

and scissors are included when applicable).<br />

Available as a KIT<br />

Includes CAL CHECK standard cuvets, cuvet<br />

cleaning cloth and scissors (when applicable)<br />

bundled in a deluxe hardshell carrying case.<br />

Just add a “C” at the end of<br />

the part #. Eg: HI 96715C<br />

HI 95<br />

Series<br />

Available<br />

WWW. HANNAINST. COM<br />

113

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!