Chapter 3: Achieving Good Recordings<br />
You must refer to other sources for general EEG information, description of proper EEG recording<br />
techniques, and for detailed discussion of EEG artifacts. A general knowledge of EEG is essential to<br />
obtaining valid data. Please read the Electrode Configuration Section of this manual before attempting<br />
EEG recording.<br />
Techniques to Obtain the Best EEG Recordings with Mindset<br />
Follow these steps to obtain the best EEG recordings when using Mindset:<br />
1. Assure proper electrode impedance values for your electrode system.<br />
2. Restrict movement of the subject while recording. Using a rolled towel to support the<br />
subject’s neck will minimize head movement.<br />
3. Watch the maximum amplitude of the input signal. If the amplitude exceeds 120 microvolts<br />
peak (or 240 microvolts peak-to-peak), the amplifiers will saturate and the resultant data is<br />
clipped and invalid.<br />
Figure 3.1 - Amplifier Saturation<br />
If you need to extend Mindset’s amplitude range, contact your distributor about the optional<br />
gain variance modules.<br />