6. ELECTRONIC TRANSMITTER (-E)<br />
6.2.8 Error messages (F4-)<br />
If the 8 bars beneath the digits start flashing, an error has occurred in the measuring transmitter/current<br />
output. Since the pointer indication is independent from the electric measuring transmitter, it may show the<br />
correct measuring value even if the transmitter is defective. Function F41 allows checking of assigned error<br />
codes.<br />
Error codes are called onto the display as follows:<br />
Description Selection Key Indication<br />
Display mode<br />
Change to setting mode Enter F1-<br />
Setting mode 3x → F4-<br />
Enter F41<br />
Enter F41 Enn<br />
Enter F41<br />
Back to display mode ↑ F1-<br />
↑ Display mode<br />
List of error messages :<br />
Code Meaning Remedy<br />
01 RAM-error Indication unit needs service<br />
02 ADC-error Indication unit needs service<br />
03 Internal EEPROM faulty Indication unit needs service<br />
04 Calibration-EEPROM faulty If EEPROM is missing insert,-otherwise order<br />
new EEPROM<br />
05 Wrong totalizer value in EEPROM Reset totalizer<br />
06 Overflow (flow rate too high) Reduce flow rate<br />
07 Internal EEPROM faulty Indication unit needs service<br />
08 Float blocking indication realized, Deactivate float-blocking-indication or run<br />
supervision time gone off<br />
Autozero function<br />
In case of error the appropriate remedy das to be taken.<br />