Brief Manual<br />

Nano Technology Systems Division<br />

LIBRA® <strong>120</strong><br />

Transmission Electron Microscope<br />

Enabling the Nano-Age World


User´s manual<br />

Nano Technology Systems Division<br />

LIBRA <strong>120</strong><br />

Operating Manual<br />

344200-0000-016


© Carl <strong>Zeiss</strong> NTS GmbH<br />

A Carl <strong>Zeiss</strong> SMT AG Company<br />

Carl-<strong>Zeiss</strong>-Str. 56<br />

73447 Oberkoch<strong>en</strong><br />

Germany<br />

This manual is designed for users (operators) who have be<strong>en</strong> trained in the use of<br />

the instrum<strong>en</strong>t by an authorized professional. Users of this instrum<strong>en</strong>t may not<br />

deviate from the instruction provided in this docum<strong>en</strong>tation.<br />

Carl <strong>Zeiss</strong> NTS GmbH holds all copyrights and rights of exploitation, including all<br />

translations of this material. The lic<strong>en</strong>se rights for the software provided with this<br />

instrum<strong>en</strong>t are noted in an agreem<strong>en</strong>t that becomes a perman<strong>en</strong>t part of the<br />

instrum<strong>en</strong>t file.<br />

No part of this docum<strong>en</strong>tation may be reproduced, edited, or provided to a third<br />

party without writt<strong>en</strong> permission from Carl <strong>Zeiss</strong> NTS GmbH.<br />

If the user modifies the instrum<strong>en</strong>t without the manufacturer´s cons<strong>en</strong>t, the<br />

manufacturer will not be liable for any damage that may result. Such modifications<br />

will also void the warranty.<br />

Carl <strong>Zeiss</strong> NTS GmbH reserves the right to terminate the user´s lic<strong>en</strong>se and sue for<br />

damages if the terms of this agreem<strong>en</strong>t are violated.<br />

Carl <strong>Zeiss</strong> NTS GmbH reserves the right to make modifications as part of the<br />

technical developm<strong>en</strong>t process.<br />

Carl <strong>Zeiss</strong> NTS GmbH<br />

A Carl <strong>Zeiss</strong> SMT AG Company<br />

Carl <strong>Zeiss</strong> Straße 56<br />

D-73447 Oberkoch<strong>en</strong><br />

GERMANY<br />

Tel: +497364/205991<br />

Fax: +497364/943226<br />

info-nts@smt.zeiss.com<br />

Carl <strong>Zeiss</strong> SMT Ltd.<br />

Coldhams Lane<br />

Cambridge CB1 3JS<br />

UK<br />

Tel: +44 1223/414166<br />

Fax: +44 1223/412776<br />

info-uk@smt.zeiss.com<br />

Carl <strong>Zeiss</strong> SMT Sarl.<br />

86, Av<strong>en</strong>ue du 18 juin 1940<br />

92500 Rueil-Malmaison<br />

France<br />

Tel: +33 1 41/39 92 10<br />

Fax: +33 1 41/39 92 29<br />

info-fr@smt.zeiss.com<br />

Carl <strong>Zeiss</strong> SMT Inc.<br />

One <strong>Zeiss</strong> Drive,Thornwood<br />

New York 10594<br />

USA<br />

Tel: +1 914 747 7700<br />

Fax: +1 914 747 7443<br />

info-usa@smt.zeiss.com


Brief Manual LIBRA <strong>120</strong><br />

1. EM System Turn-ON<br />

• Push the gre<strong>en</strong> button (On) at the console.<br />

• Start the EM PC.<br />

• Click on the icon WinTEM on the Desktop.<br />

• Type in user ID and password.<br />

2. Check Vacuum Readiness<br />

− Column, Gun, and Camera chamber<br />

are gre<strong>en</strong>.<br />

3. Turn On High T<strong>en</strong>sion<br />

• Op<strong>en</strong> Gun tab sheet.<br />

• Click on the icon HT On.<br />

− HT is ramping up to the selected<br />

high voltage step.<br />

4. Turn On Filam<strong>en</strong>t<br />

• Click on the icon Fil On.<br />

− The filam<strong>en</strong>t is ramping up according to the<br />

gun profile.<br />

• Adjust emission step until the emission curr<strong>en</strong>t<br />

is ≥ 1 µA.<br />

1


5. Check Status of the Specim<strong>en</strong> Holder<br />

• The holder must be inserted, otherwise the<br />

gun blanker cannot be op<strong>en</strong>ed.<br />

6. Op<strong>en</strong> Gun and Column Valve<br />

• Click on Column Valve Op<strong>en</strong> (if closed).<br />

− If not possible, remove holder from the airlock<br />

and insert it again.<br />

7. Check Status of the Beam Blanker<br />

• Click on the beam blanker icon that the frame<br />

around the icon is gone ⇒ Beam Blanker Off.<br />

− The status is also diplayed in the status bar<br />

and can be changed there.<br />

• Check also that no other device is blanking the<br />

beam externally.<br />

8. Retract Specim<strong>en</strong> Holder<br />

• Retract holder untill it stops (1) and turn it a<br />

few degrees to the left (2) that it stays retracted.<br />

− Make sure that WinTEM still recognizes the<br />

holder to be inserted.<br />

2


9. Check Mag and Ill Angle<br />

• Set mag to 10k and illumination angle to 0.5 mrad.<br />

☞<br />

NOTICE:<br />

If there is fluor<strong>en</strong>s<strong>en</strong>ce on the scre<strong>en</strong>, continue with point 12.<br />

If there is no fluor<strong>en</strong>s<strong>en</strong>ce on the scre<strong>en</strong>, continue with point 10.<br />

10. Retract Apertures<br />

• Retract spectrometer slit aperture.<br />

− Mostly the beam is visible on the scre<strong>en</strong>.<br />

− If so, continue with 11.<br />

• Retract objective aperture.<br />

• Retract cond<strong>en</strong>ser aperture.<br />

11. Calibrate Deflecting Systems<br />

• Select Img Shft and calibrate it with R in the<br />

right lower corner of the nav. box.<br />

• Select Calibrate All , if no fluores<strong>en</strong>ce is<br />

visible on the scre<strong>en</strong>.<br />

This way is recomm<strong>en</strong>ded, if the status of the<br />

LIBRA <strong>120</strong> is not known.<br />

3


12. Turn AIS Mode to Off<br />

• Click on AIS Off in the tab sheet TEM.<br />

− Display in the status bar.<br />

• Set brightness to 0.5 mrad.<br />

13. Select Magnification<br />

• Select mag 31.5 k with knob (3).<br />

M8<br />

M7<br />

M6<br />

M5<br />

2<br />

3<br />

14. Check Spot Position<br />

• Switch to Spot mode with button (20).<br />

• Calibrate spot with button Cal (25).<br />

• C<strong>en</strong>ter spot to index point with knobs (1,6).<br />

• Switch back to TEM illumination with button (20).<br />

M4<br />

M3<br />

M2<br />

M1<br />

20<br />

25<br />

5<br />

43<br />

30<br />

31<br />

1 6<br />

15. Insert Cond<strong>en</strong>ser Aperture<br />

• Turn cond<strong>en</strong>ser aperture holder to the 3. clickstop<br />

and c<strong>en</strong>ter the c<strong>en</strong>tral AIS aperture (37.5 µm).<br />

•<br />

4


16. Check Spectrum Position<br />

• Switch to EELS mode with button (23).<br />

− If the spectrum tip is not visible on the large<br />

scre<strong>en</strong>, start the magnetic reset as described<br />

in point 17.<br />

• Shift spectrum tip to the index point<br />

of the scre<strong>en</strong> with knobs (1, 6).<br />

• Switch to ESI mode with button (23).<br />

M4<br />

M3<br />

M2<br />

M1<br />

23<br />

43<br />

30 31<br />

1 6<br />

Index point<br />

17. Start Magnetic Reset<br />

• Click on Start Magnetic Reset to formate the<br />

OMEGA filter.<br />

− Only necessary, if spectrum is not visible<br />

on the scre<strong>en</strong>.<br />

− This can take 1-2minutes.<br />

18. Turn AIS Mode to Auto<br />

• Click on AIS Auto in the TEM tab sheet.<br />

− Display in the status bar.<br />

19. C<strong>en</strong>ter Spectrometer Entrance Aperture<br />

• Insert aperture (2. clickstop) and c<strong>en</strong>ter it to<br />

the large scre<strong>en</strong>.<br />

5


20. Adjust Spot<br />

• Increase magnification to 100k with knob (3)<br />

• Switch to Spot mode with button (20).<br />

• Increase spot size to 100 nm with knob (5).<br />

• Push Cal (25).<br />

• Shift spot to the index point with knobs (1, 6).<br />

M4<br />

M3<br />

M2<br />

M1<br />

20<br />

25<br />

5<br />

43<br />

30<br />

31<br />

1 6<br />

• Rec<strong>en</strong>ter spot to index point with knobs (1, 6).<br />

• Switch back to TEM mode with button (20)..<br />

21. C<strong>en</strong>ter Cond<strong>en</strong>ser Aperture<br />

• Click on AIS Off in the tab sheet TEM.<br />

− Display in the status bar.<br />

• C<strong>en</strong>ter cond<strong>en</strong>ser aperture (37.5 µm) to the<br />

large scre<strong>en</strong> mechanically.<br />

22. Insert Specim<strong>en</strong> Holder<br />

• Insert objective aperture (1. clickstop).<br />

− see point 25 for adjustm<strong>en</strong>t.<br />

• Lower mag to 50 k.<br />

• Move holder back into the column and look<br />

for a specim<strong>en</strong> detail.<br />

• C<strong>en</strong>ter detail to the index point.<br />

23. Adjust Euc<strong>en</strong>tric Height<br />

• Switch on Foc Aid with button (26).<br />

• Calibrate the objective l<strong>en</strong>s with button (25).<br />

− The sample is moving forth and back, if it is<br />

not in the euc<strong>en</strong>tric plane.<br />

• Adjust Z with the focus knob (4).<br />

M4<br />

M3<br />

M2<br />

M1<br />

25<br />

22<br />

26<br />

4<br />

5<br />

• Turn off Foc Aid with button (26).<br />

6


24. Check Illumination Tilt<br />

• Move a specim<strong>en</strong> detail to the scre<strong>en</strong> c<strong>en</strong>ter<br />

with the joystick.<br />

• Turn on the wobbler Objective in the group box<br />

Wobbler.<br />

• Minimize the oscillating movem<strong>en</strong>t of the sample<br />

with knobs (1, 6).<br />

• Turn off the wobbler with Off.<br />

25. Check Beam Alignm<strong>en</strong>t<br />

• Select Diff mode with button (22).<br />

• Set illumination angle (brightness)<br />

to 0.06 mrad (60 µrad) with knob (5).<br />

• Adjust diffraction spot to index point with<br />

knobs (1) and (6).<br />

• Activate Gun Tilt in the group box Alignm<strong>en</strong>t.<br />

− Gun Tilt is displayed on the hardpanel (43).<br />

• Increase illumination angle (brightness) to<br />

1.6 mrad with knob (5).<br />

• Shift diffraction spot to the index point with<br />

knobs (1) and (6).<br />

• Save settings with Store Gun Alignm<strong>en</strong>t.<br />

26. Check Diffraction Spot<br />

• Select Diff mode (22) and set camera l<strong>en</strong>gth<br />

to 720 mm with knob (3) on the left hardpanel.<br />

• Shift diffraction spot to the index point with<br />

knobs (1, 6).<br />

• Insert an objective aperture and c<strong>en</strong>ter it to<br />

the diffraction spot.<br />

• Switch back to IMG mode with button (22).<br />

M4<br />

M3<br />

M2<br />

M1<br />

J1<br />

22<br />

43<br />

30<br />

31<br />

1 6<br />

7


27. Turn AIS Mode to Auto<br />

• Click on AIS Auto.<br />

− Display Auto in the status bar.<br />

If the AIS alignm<strong>en</strong>t was done and saved earlier, there is nothing to do on AIS.<br />

☞<br />

NOTICE:<br />

If an AIS alignm<strong>en</strong>t is necessary, go to chapter AIS Alignm<strong>en</strong>t in the ext<strong>en</strong>ded manual.<br />

8


Enabling the Nano-Age World<br />

Carl <strong>Zeiss</strong> NTS GmbH<br />

A Carl <strong>Zeiss</strong> SMT AG Company<br />

Carl <strong>Zeiss</strong> Straße 56<br />

D-73447 Oberkoch<strong>en</strong><br />

GERMANY<br />

Tel: +497364/205991<br />

Fax: +497364/943226<br />

info-nts@smt.zeiss.com<br />

Carl <strong>Zeiss</strong> SMT Ltd.<br />

Coldhams Lane<br />

Cambridge CB1 3JS<br />

UK<br />

Tel: +44 1223/414166<br />

Fax: +44 1223/412776<br />

info-uk@smt.zeiss.com<br />

Carl <strong>Zeiss</strong> SMT Inc.<br />

One <strong>Zeiss</strong> Drive,Thornwood<br />

New York 10594<br />

USA<br />

Tel: +1 914 747 7700<br />

Fax: +1 914 747 7443<br />

info-usa@smt.zeiss.com<br />

Carl <strong>Zeiss</strong> SMT Sarl<br />

86, av<strong>en</strong>ue du 18 juin 1940<br />

Rueil-Malmaision<br />

FRANCE<br />

Tel: +33 1 41399210<br />

Fax: +33 1 41399229<br />

info-fr@smt.zeiss.com<br />

Plus a worldwide network<br />

of authorised distributors<br />

www.smt.zeiss.com<br />

Due to policy of continuous<br />

developm<strong>en</strong>t, we reserve the<br />

right to change specifications<br />

without notice.<br />

© Carl <strong>Zeiss</strong> SMT, Oberkoch<strong>en</strong>

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!