Cleaning the Valve Plates<br />

In case that the leaktest is not passed, a very<br />

common reason is dirt on the rubber plates of<br />

the pump valves.<br />

This problem can easily be solved by cleaning<br />

the valve plates:<br />

- dismount the pump from the Gas 1000 housing<br />

- remove the valve chambers from the pump<br />

D 0000-16<br />